ZAP-X Radiosurgery System

Published September 21, 2022 · Recall initiated: July 5, 2022

Summary

FDA records medical device Class II recall Z-1742-2022, published September 21, 2022 and initiated July 5, 2022. Recalling firm: Zap Surgical Systems; Status: Ongoing; Reported quantity: 8; Lot: UDI-DI: 00860183001504, Serial Numbers, ZUC17152, ZUC17153, ZUC18155, ZUC20157, ZUC20158, ZUC20156, ZUC21161, ZUC21162, using Treatment Delivery Software Version: 1.8.53. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Software issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these cases the transitional alignment algorithm may incorrectly calculate the new treatment table offset values, leading to an incorrect position for the subsequent treatment isocenter.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Zap Surgical Systems
Distributor
Worldwide distribution - US Nationwide distribution in the states of AZ, MD, FL, CO and the countries of DEU, CHE, JPN.
Lot
UDI-DI: 00860183001504, Serial Numbers, ZUC17152, ZUC17153, ZUC18155, ZUC20157, ZUC20158, ZUC20156, ZUC21161, ZUC21162, using Treatment Delivery Software Version: 1.8.53
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ZAP-X Radiosurgery System recalled?

Yes. FDA published recall Z-1742-2022 on September 21, 2022. Reported hazard: Software issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these cases the transitional alignment algorithm may incorrectly calculate the new treatment table offset values, leading to an incorrect position for the subsequent treatment isocenter.

What should I do if I have this product?

Voluntary: Firm initiated · Letter