BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tubes, 4.5 mL, 13 x 75 mm, Mfd: Becton Dickinson and Company Product Usage: The BD Vacutainer Safety Coagulation tube is an evacuated blood collection Tube that provides means of collecting, transporting and processing blood in a closed tube. The buffered sodium citrate additive provides an anti-coagulated specimen that may be used for clinical laboratory coagulation assays and determinations.
Published April 12, 2017 · Recall initiated: March 16, 2017
Summary
FDA records medical device Class II recall Z-1746-2017, published April 12, 2017 and initiated March 16, 2017. Recalling firm: Becton Dickinson & Company; Status: Terminated; Reported quantity: 13,064,500 tubes (non-expired); Lot: Catalog Number: 367716 - Lot numbers: 5251673 with expiry 3/31/2017; 6064542 with expiry 9/30/17; 6187520 with expiry 1/31/2018 and 6253669 with expiy 3/31/18 Catalog Number: 369714 - Lot numbers: 5272777 with expiry 4/30/17; 5336909 with expiry 6/30/17; 5357632 with expiry 7/31/17; 6040953 and 6064634 both with expiry 9/30/17; 6090572 with expiry 10/31/17; 6124932 with expiry 11/30/17; 6173983 with expiry 12/31/17; 6187600 with expiry 1/31/18 and 6216656 with expiry 2/28/18 Lot numbers 6064542, 6187520 and 6253669 - limited to distribution outside the U.S. market. UDI for lot 6253669 is DI (01)30382903677161, PI (17)180331(10)6253669(30)0100 and lot 6216656 is DI (01)30382903697145, PI (17)180228(10)6216656(30)0100; for all other lots UDI is not applicable based on manufacture date. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
BD is initiating this product correction of multiple lots of BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tube (13x75 mm x 4.5 mL). A limited number of tubes in these lots may exhibit stopper pullout from the tube as the user removes the needle from the stopper following specimen collection.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Becton Dickinson & Company
- Distributor
- Worldwide - US Nationwide Distribution
- Lot
- Catalog Number: 367716 - Lot numbers: 5251673 with expiry 3/31/2017; 6064542 with expiry 9/30/17; 6187520 with expiry 1/31/2018 and 6253669 with expiy 3/31/18 Catalog Number: 369714 - Lot numbers: 5272777 with expiry 4/30/17; 5336909 with expiry 6/30/17; 5357632 with expiry 7/31/17; 6040953 and 6064634 both with expiry 9/30/17; 6090572 with expiry 10/31/17; 6124932 with expiry 11/30/17; 6173983 with expiry 12/31/17; 6187600 with expiry 1/31/18 and 6216656 with expiry 2/28/18 Lot numbers 6064542, 6187520 and 6253669 - limited to distribution outside the U.S. market. UDI for lot 6253669 is DI (01)30382903677161, PI (17)180331(10)6253669(30)0100 and lot 6216656 is DI (01)30382903697145, PI (17)180228(10)6216656(30)0100; for all other lots UDI is not applicable based on manufacture date.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABD®Spinal Tray with BD®Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDABD®Spinal Tray with BD®Quincke Needle 26 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDABD®Spinal Tray with BD®Whitacre Needle 22 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDABD®Spinal Tray with BD®Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 2 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDACovidien Curity Saline Dressing 8' x 4 (20.3cm x10.2 cm), Sterile Item Code: 3606 Curity sodium chlorideApril 12, 2017
- FDASiemens ADVIA Centaur BR assay Cat No. 03896216, SMN 10334837 (5-pack/250 tests) BR Assay for CA 27.29 TheApril 12, 2017
Questions about this recall
Was BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tubes, 4.5 mL, 13 x 75 mm, Mfd: Becton Dickinson and Company Product Usage: The BD Vacutainer Safety Coagulation tube is an evacuated blood collection Tube that provides means of collecting, transporting and processing blood in a closed tube. The buffered sodium citrate additive provides an anti-coagulated specimen that may be used for clinical laboratory coagulation assays and determinations. recalled?
Yes. FDA published recall Z-1746-2017 on April 12, 2017. Reported hazard: BD is initiating this product correction of multiple lots of BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tube (13x75 mm x 4.5 mL). A limited number of tubes in these lots may exhibit stopper pullout from the tube as the user removes the needle from the stopper following specimen collection.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit