Clinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum covering 38 commonly used clinical chemistry tests.
Published May 15, 2024 · Recall initiated: March 5, 2024
Summary
FDA records medical device Class II recall Z-1759-2024, published May 15, 2024 and initiated March 5, 2024. Recalling firm: Randox Laboratories Ltd.; Status: Ongoing; Reported quantity: 80,366 units (US: 242 units; OUS: 80,094 units); Lot: Catalogue No. CAL2351 & CAL10388; GTIN: 05055273200966; Lot No. 1260UE, 1262UE, 1295UE, 1297UE, 1298UE, 1315UE, 1325UE, 1326UE, 1295UE, & 1325UE. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Randox Laboratories has identified that Inorganic Phosphate in Calibration Serum Level 3, CAL2351 & CAL10388, is running with a negative bias on RX Series instruments compared to other methods.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Randox Laboratories Ltd.
- Distributor
- Domestic: Nationwide Distribution.
- Lot
- Catalogue No. CAL2351 & CAL10388; GTIN: 05055273200966; Lot No. 1260UE, 1262UE, 1295UE, 1297UE, 1298UE, 1315UE, 1325UE, 1326UE, 1295UE, & 1325UE.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Ireland
Related recalls
- FDARX Series CYSC- Cystatin C Reagent -Intended for in vitro diagnostic (IVD) use as Randox Cystatin C assay onJuly 31, 2024
- FDAEvidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drugMay 15, 2024
- FDAPERLA TL 25D SCREW REF: TLF-DS 45 25-S TLF-DS 45 30-S TLF-DS 45 35-S TLF-DS 45 40-S TLF-DS 45 45-S TLF-DS 55May 15, 2024
- FDADURAFLOW 2, 24CM STRAIGHT BASIC KIT -Indicated for use in attaining Long-Term vascular access for HemodialysisMay 15, 2024
- FDAMicroalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567April 24, 2024
- FDAMicroalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426April 24, 2024
Questions about this recall
Was Clinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum covering 38 commonly used clinical chemistry tests. recalled?
Yes. FDA published recall Z-1759-2024 on May 15, 2024. Reported hazard: Randox Laboratories has identified that Inorganic Phosphate in Calibration Serum Level 3, CAL2351 & CAL10388, is running with a negative bias on RX Series instruments compared to other methods.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail