GE Healthcare Allia IGS 5 Pulse, Interventional fluoroscopic x-ray systems
Published May 15, 2024 · Recall initiated: March 18, 2024
Summary
FDA records medical device Class II recall Z-1782-2024, published May 15, 2024 and initiated March 18, 2024. Recalling firm: GE Medical Systems, SCS; Status: Ongoing; Reported quantity: 1 unit; Lot: UDI/DI 00195278719263, Serial Numbers: M3-23-117. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE HealthCare has become aware of the potential that adhesive material might not have been applied to some bolts of certain Innova IGS, Discovery IGS, Interact IGS, Allia IGS and Allia Pulse IGS products. This could result in a potential fall of the Detector/Elevator or X-ray tube of the frontal Gantry assembly.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Medical Systems, SCS
- Distributor
- Worldwide distribution - US Nationwide in the states of CA, FL, GA, IN, MA, NC, OK, SC, TN, TX, VA and the countries of Armenia, China, France, Hungary, India, Italy, Japan, Kazakhstan, Lithuania, Mexico, Russia, Spain, Taiwan, Turkey, Vietnam
- Lot
- UDI/DI 00195278719263, Serial Numbers: M3-23-117
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- France
Related recalls
- FDAAW Server 2.0, AW Server 3.0, AW Server 3.1, AW Server 3.2October 15, 2025
- FDAGE Healthcare Innova IGS 6, Interventional fluoroscopic x-ray systemsMay 15, 2024
- FDAGE Healthcare Allia IGS 5 Pulse, Interventional fluoroscopic x-ray systemsMay 15, 2024
- FDAGE Healthcare Interact Discovery RT, Interventional fluoroscopic x-ray systemsMay 15, 2024
- FDAGE Healthcare Allia IGS 7 OR, Interventional fluoroscopic x-ray systemsMay 15, 2024
- FDAGE Healthcare Allia IGS 7, Interventional fluoroscopic x-ray systemsMay 15, 2024
Questions about this recall
Was GE Healthcare Allia IGS 5 Pulse, Interventional fluoroscopic x-ray systems recalled?
Yes. FDA published recall Z-1782-2024 on May 15, 2024. Reported hazard: GE HealthCare has become aware of the potential that adhesive material might not have been applied to some bolts of certain Innova IGS, Discovery IGS, Interact IGS, Allia IGS and Allia Pulse IGS products. This could result in a potential fall of the Detector/Elevator or X-ray tube of the frontal Gantry assembly.
What should I do if I have this product?
Voluntary: Firm initiated · Letter