Product Name: MR 7700; Model Numbers: (1) 782120, (2) 782153;
Published May 28, 2025 · Recall initiated: May 13, 2025
Summary
FDA records medical device Class II recall Z-1805-2025, published May 28, 2025 and initiated May 13, 2025. Recalling firm: Philips North America Llc; Status: Ongoing; Reported quantity: 72 units (6 US, 66 OUS); Lot: Model Numbers: (1) 782120, (2) 782153; UDI-DI: (1) 00884838104112, (2) 00884838112858; Serial Numbers: (1) 65000, 65001, 65002, 65003, 65004, 65005, 65007, 65008, 65009, 65011, 65012, 65013, 65014, 65017, 65018, 65019, 65020, 65021, 65023, 65024, 65025, 65026, 65027, 65028, 65029, 65030, 65031, 65032, 65034, 65035, 65037, 65038, 65039, 65040, 65041, 65042, 65050, 65051, 65056, 65057, 65059, 65060, 65061, 65063, 65064, 65065, 65066, (2) 68000, 68001, 68006, 68007, 68008, 68009, 68010, 68011, 68012, 68013, 68014, 68015, 68016, 68017, 68018, 68020, 68023, 68024, 68025, 68028, 68029, 68030, 68031, 68034, 68037. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips North America Llc
- Distributor
- Worldwide distribution - US Nationwide and the countries of Andorra, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, C®te D'Ivoire, Cyprus, Czech Republic, Denmark, Finland, France, French Guiana, Georgia, Germany, Greece, Guadeloupe, Hong Kong, Iceland, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Latvia, Lebanon, Libya, Lithuania, Malaysia, Malta, Martinique, Mexico, Moldova, Monaco, Netherlands, New Zealand, Norway, Palestine, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Serbia, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Kingdom, U.A.E.
- Lot
- Model Numbers: (1) 782120, (2) 782153; UDI-DI: (1) 00884838104112, (2) 00884838112858; Serial Numbers: (1) 65000, 65001, 65002, 65003, 65004, 65005, 65007, 65008, 65009, 65011, 65012, 65013, 65014, 65017, 65018, 65019, 65020, 65021, 65023, 65024, 65025, 65026, 65027, 65028, 65029, 65030, 65031, 65032, 65034, 65035, 65037, 65038, 65039, 65040, 65041, 65042, 65050, 65051, 65056, 65057, 65059, 65060, 65061, 65063, 65064, 65065, 65066, (2) 68000, 68001, 68006, 68007, 68008, 68009, 68010, 68011, 68012, 68013, 68014, 68015, 68016, 68017, 68018, 68020, 68023, 68024, 68025, 68028, 68029, 68030, 68031, 68034, 68037;
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333October 7, 2026
- FDAAchieva 1.5T, Achieva 1.5T Conversion and Achieva 1.5T Initial system; Model Numbers (REF): 1. 781196 (AchievaSeptember 30, 2026
- FDAIngenia 3.0T and Ingenia 3.0T CX; Model Numbers (REF): 1. 781342(3.0T), 2. 781377(3.0T), 3. 782103(3.0T), 4.September 30, 2026
- FDAIngenia 1.5T, Ingenia 1.5T CX and Ingenia 1.5T S; Model Numbers (REF): 1. 781315(1.5T), 2. 781341(1.5T), 3.September 30, 2026
- FDAProduct Name: Ingenia 3.0T CX; Model Number: 781271;May 28, 2025
- FDATRUE 3D VIEWER SYSTEM Model Number EP-00001-XX Software versions 1.0, 1.25, and 1.35 The t3D-Viewer isMay 28, 2025
Questions about this recall
Was Product Name: MR 7700; Model Numbers: (1) 782120, (2) 782153; recalled?
Yes. FDA published recall Z-1805-2025 on May 28, 2025. Reported hazard: Potential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.
What should I do if I have this product?
Voluntary: Firm initiated · Letter