VITROS Immunodiagnostic Products Free T3 Calibrators-IVD are for use with Free T3 Reagent Pack for the quantitative measurement of free triiodothyronine (FT3) in human serum and plasma (EDTA or heparin). Product Code: 111 2820
Published May 22, 2024 · Recall initiated: April 9, 2024
Summary
FDA records medical device Class II recall Z-1819-2024, published May 22, 2024 and initiated April 9, 2024. Recalling firm: QUIDELORTHO; Status: Ongoing; Reported quantity: N/A; Lot: UDI-DI: 10758750008438 Lot Numbers: 3042 Expiry: 09-Apr-2024; 3052 Expiry: 02-May-2024; 3061 Expiry: 06-Jun-2024; 3080 Expiry: 02-May-2024; 3090 Expiry: 14-Jun-2024; 3100 Expiry: 29-Jul-2024; 3110 Expiry: 13-Aug-2024; 3126 Expiry: 04-Sep-2024. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Free T3 Calibrators used with T3 Reagent Packs may produce positively biased results for patient and quality control samples. The positive bias may cause erroneous patient results or a delay in testing that may impact patient management, clinical assessment, and the results of other thyroid function tests.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- QUIDELORTHO
- Distributor
- Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore , Spain, Sweden, The Netherlands, United Kingdom.
- Lot
- UDI-DI: 10758750008438 Lot Numbers: 3042 Expiry: 09-Apr-2024; 3052 Expiry: 02-May-2024; 3061 Expiry: 06-Jun-2024; 3080 Expiry: 02-May-2024; 3090 Expiry: 14-Jun-2024; 3100 Expiry: 29-Jul-2024; 3110 Expiry: 13-Aug-2024; 3126 Expiry: 04-Sep-2024
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVITROS Immunodiagnostic Products Free T3 Reagent Pack-IVD is for the quantitative measurement of freeMay 22, 2024
- FDAHeartSine samaritan PAD (Public Access Defibrillator), Type: SAM 350P, SAM 360P, SAM 450PMay 22, 2024
- FDASpectrum IQ Infusion pump, Product Code 357009May 22, 2024
- FDAAPTUS 2.8 TriLock Screw 22mm, HD7, 1/Pkg. Metal bone fixation fastener used with the APTUS fixation system.May 22, 2024
- FDALASER Application Instrument, 23 cm, REF: 8694; LEUNIG-GREVERS LASER Application, REF: 461000May 22, 2024
- FDANovalung sensor box is a component of the Novalung Console (F30000162)-Indicated for long-term (> 6 hours)May 22, 2024
Questions about this recall
Was VITROS Immunodiagnostic Products Free T3 Calibrators-IVD are for use with Free T3 Reagent Pack for the quantitative measurement of free triiodothyronine (FT3) in human serum and plasma (EDTA or heparin). Product Code: 111 2820 recalled?
Yes. FDA published recall Z-1819-2024 on May 22, 2024. Reported hazard: Free T3 Calibrators used with T3 Reagent Packs may produce positively biased results for patient and quality control samples. The positive bias may cause erroneous patient results or a delay in testing that may impact patient management, clinical assessment, and the results of other thyroid function tests.
What should I do if I have this product?
Voluntary: Firm initiated · Letter