Back to search
FDAFDA · Food and drugs·Recall number Z-1828-2013

Medline Med-Checks BI Test packs with Instant Readout Integrator, SteriTec Product Code PL 200, Distributor Product Code MDS200500IR SteriTec Steam Biological Test Packs with Instant Readout Integrator Card: The Steam Biological Test Pack with Instant Readout Integrator is a steam sterilization monitor designed specifically for biological testing of pre-vacuum steam sterilizers operating at 132®C (270®F) for 3 minutes or longer. The Instant Readout Integrator Card gives the operator a prediction of the biological test outcome

Published August 7, 2013 · Recall initiated: June 10, 2013

Summary

FDA records medical device Class II recall Z-1828-2013, published August 7, 2013 and initiated June 10, 2013. Recalling firm: Steritec Products Mfc. Co., Inc.; Status: Terminated; Reported quantity: 3535 units; Lot: Affected Biological Indicator Lot No.:130401. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Mesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest Steam and Smart-Read Biological Indicators (BI) because the readout time is not in specification with the label claims.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Steritec Products Mfc. Co., Inc.
Distributor
Worldwide distribution: US (Nationwide) and countries of: Argentina, Australia, Canada, British Virgin Island, Czech Republic, Italy, and Turkey.
Lot
Affected Biological Indicator Lot No.:130401
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Medline Med-Checks BI Test packs with Instant Readout Integrator, SteriTec Product Code PL 200, Distributor Product Code MDS200500IR SteriTec Steam Biological Test Packs with Instant Readout Integrator Card: The Steam Biological Test Pack with Instant Readout Integrator is a steam sterilization monitor designed specifically for biological testing of pre-vacuum steam sterilizers operating at 132®C (270®F) for 3 minutes or longer. The Instant Readout Integrator Card gives the operator a prediction of the biological test outcome recalled?

Yes. FDA published recall Z-1828-2013 on August 7, 2013. Reported hazard: Mesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest Steam and Smart-Read Biological Indicators (BI) because the readout time is not in specification with the label claims.

What should I do if I have this product?

Voluntary: Firm initiated · Letter