The product is Pathfinder SL, version 1.6.0.3057 software used with the Spacelabs Healthcare Pathfinder SL Holter Analyzer, version 1.6.0. Product Usage: The Holter Analyzer analyzes information collected from ambulatory electrocardiogram recorders worn by patients and used to assist physicians in diagnosis and patient monitoring. In addition, incorrect QTc values are reported after moving the cursor. QTc values are timing events between labeled points of interest in the ECG signal that the physician can mark with the cursor. The software is used to edit a family of events and upon conclusion the software crashes with an error message.
Published June 27, 2012 · Recall initiated: May 11, 2012
Summary
FDA records medical device Class II recall Z-1831-2012, published June 27, 2012 and initiated May 11, 2012. Recalling firm: Spacelabs Healthcare, Llc; Status: Terminated; Reported quantity: 114; Lot: Version 1.6.0. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm received reports of the Pathfinder LS software crashing after editing a batch of Holter Analyzer records. The software will stop with an error message requiring a restart. In addition, QTc values will not update after moving the marker.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Spacelabs Healthcare, Llc
- Distributor
- Worldwide Distribution - US (nationwide) including the states of Nevada and Vermont, and the countries of AUSTRALIA, AUSTRIA, CANADA, CYPRUS, GERMANY, ICELAND, ISRAEL, POLAND, ROMANIA, SOUTH AFRICA, SWITZERLAND, THAILAND, TURKEY and UNITED KINGDOM.
- Lot
- Version 1.6.0.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASpacelabs Healthcare G2 Clinical Access Mislabeled CDs for PN: 063-1829-11 Rev A and Update Kit, PN:April 2, 2014
- FDAARIATELE TELEMETRY TRANSMITTER, Model 96281, with SpO2 Option C. Model 96281 is intended for use with eitherFebruary 5, 2014
- FDASpacelabs Healthcare qube Compact Patient Monitor, Model 91390. It is a patient monitor with arrhythmiaNovember 27, 2013
- FDASpacelabs Healthcare elance Vital Signs Monitor, Model 93300, with Option S. It is a patient monitor withNovember 20, 2013
- FDAPhilips Multi Diagnost Eleva II with swivel cable Product Codes; 708032, 708036 Multi Diagnost Eleva withJune 27, 2012
- FDASARNS HIGH-FLOW AORTIC ARCH CANNULA: 5.2 MM (16 FR) 00 WITH 3/8" CONNECTOR, 7.5" (19 CM) LONG. The Sarns HighJune 27, 2012
Questions about this recall
Was The product is Pathfinder SL, version 1.6.0.3057 software used with the Spacelabs Healthcare Pathfinder SL Holter Analyzer, version 1.6.0. Product Usage: The Holter Analyzer analyzes information collected from ambulatory electrocardiogram recorders worn by patients and used to assist physicians in diagnosis and patient monitoring. In addition, incorrect QTc values are reported after moving the cursor. QTc values are timing events between labeled points of interest in the ECG signal that the physician can mark with the cursor. The software is used to edit a family of events and upon conclusion the software crashes with an error message. recalled?
Yes. FDA published recall Z-1831-2012 on June 27, 2012. Reported hazard: The firm received reports of the Pathfinder LS software crashing after editing a batch of Holter Analyzer records. The software will stop with an error message requiring a restart. In addition, QTc values will not update after moving the marker.
What should I do if I have this product?
Voluntary: Firm initiated · Letter