NIPRO Safe Touch Safety Scalp Vein Set

Published June 25, 2014 · Recall initiated: May 4, 2013

Summary

FDA records medical device Class II recall Z-1834-2014, published June 25, 2014 and initiated May 4, 2013. Recalling firm: Nipro Medical Corporation; Status: Terminated; Reported quantity: 176,000 pieces; Lot: Product Code: PR+25G19, Lots #s 12H31H and 12I14H; Product Code: SPR+21G19, Lots #s 12I24 and 12J11; Product Code: PR-25G19, Lot # 12H27H; and PR-23G19, Lot # 12I12H. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a possibility of a crack in the CP luer connector. The position of the crack is at the connection site of the CP luer connector and the tubing. The crack could cause or contribute to leakage of fluids.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Nipro Medical Corporation
Distributor
Worldwide Distribution - USA and the countries of Canada, Bolivia, Chile, Uruguay, Belize, Dominica, Honduras, Jamaica, Paraguay and Venezuela.
Lot
Product Code: PR+25G19, Lots #s 12H31H and 12I14H; Product Code: SPR+21G19, Lots #s 12I24 and 12J11; Product Code: PR-25G19, Lot # 12H27H; and PR-23G19, Lot # 12I12H.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was NIPRO Safe Touch Safety Scalp Vein Set recalled?

Yes. FDA published recall Z-1834-2014 on June 25, 2014. Reported hazard: There is a possibility of a crack in the CP luer connector. The position of the crack is at the connection site of the CP luer connector and the tubing. The crack could cause or contribute to leakage of fluids.

What should I do if I have this product?

Voluntary: Firm initiated · Letter