Back to search
FDAFDA · Food and drugs·Recall number Z-1834-2024

Reagent Test Strips for Blood, Urobilinogen, Bilirubin, Protein, Nitrite, Ketones, Ascorbic Acid, Glucose, pH, Specific Gravity and Leucocytes in Urine by the Dip and Read Technique.

Published May 22, 2024 · Recall initiated: April 4, 2024

Summary

FDA records medical device Class II recall Z-1834-2024, published May 22, 2024 and initiated April 4, 2024. Recalling firm: Macherey Nagel Gmbh & Co. Kg; Status: Ongoing; Reported quantity: 5,200 packages; Lot: Product #102-2285, UDI/DI: 00304040001479, LOT 65402. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The reason for the recall is one of the test pads, for the parameter leukocytes, was missing on some of the test strips. The issue was identified by a consumer by way of a Henry Schein Inc. (HSI) customer complaint. This problem can potentially result in the possibility that the parameter leucocytes (a non-specific marker for urinary track infections and an indicator for inflammatory renal processes) is interpreted instrumentally as a false negative. The following HSI item number and lot number were affected: Urispec® 11-Way, #102-2285, LOT 65402.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Macherey Nagel Gmbh & Co. Kg
Distributor
US Nationwide distribution in the states of CO, FL, IN, NV, TX.
Lot
Product #102-2285, UDI/DI: 00304040001479, LOT 65402.
Risk classification
Class II
Area (FDA)
device
Country
Germany

Related recalls

Questions about this recall

Was Reagent Test Strips for Blood, Urobilinogen, Bilirubin, Protein, Nitrite, Ketones, Ascorbic Acid, Glucose, pH, Specific Gravity and Leucocytes in Urine by the Dip and Read Technique. recalled?

Yes. FDA published recall Z-1834-2024 on May 22, 2024. Reported hazard: The reason for the recall is one of the test pads, for the parameter leukocytes, was missing on some of the test strips. The issue was identified by a consumer by way of a Henry Schein Inc. (HSI) customer complaint. This problem can potentially result in the possibility that the parameter leucocytes (a non-specific marker for urinary track infections and an indicator for inflammatory renal processes) is interpreted instrumentally as a false negative. The following HSI item number and lot number were affected: Urispec® 11-Way, #102-2285, LOT 65402.

What should I do if I have this product?

Voluntary: Firm initiated · Letter