Achieva 1.5T Initial system Model Number (REF): 781178
Published May 29, 2024 · Recall initiated: May 3, 2024
Summary
FDA records medical device Class II recall Z-1838-2024, published May 29, 2024 and initiated May 3, 2024. Recalling firm: Philips North America Llc; Status: Ongoing; Reported quantity: 2 units; Lot: UDI-DI: N/A Serial Numbers: 20282, 20303. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips North America Llc
- Distributor
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
- Lot
- UDI-DI: N/A Serial Numbers: 20282, 20303
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333October 7, 2026
- FDAAchieva 1.5T, Achieva 1.5T Conversion and Achieva 1.5T Initial system; Model Numbers (REF): 1. 781196 (AchievaSeptember 30, 2026
- FDAIngenia 3.0T and Ingenia 3.0T CX; Model Numbers (REF): 1. 781342(3.0T), 2. 781377(3.0T), 3. 782103(3.0T), 4.September 30, 2026
- FDAIngenia 1.5T, Ingenia 1.5T CX and Ingenia 1.5T S; Model Numbers (REF): 1. 781315(1.5T), 2. 781341(1.5T), 3.September 30, 2026
- FDAAchieva 1.5T Conversion- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as aMay 29, 2024
- FDAAchieva 3.0T Model Number (REF): (1) 781345; (2) 781344; (3) 781278; (4) 781277; (5) 781177May 29, 2024
Questions about this recall
Was Achieva 1.5T Initial system Model Number (REF): 781178 recalled?
Yes. FDA published recall Z-1838-2024 on May 29, 2024. Reported hazard: The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
What should I do if I have this product?
Voluntary: Firm initiated · Letter