Achieva 3.0T for PET -Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): 781477

Published May 29, 2024 · Recall initiated: May 3, 2024

Summary

FDA records medical device Class II recall Z-1840-2024, published May 29, 2024 and initiated May 3, 2024. Recalling firm: Philips North America Llc; Status: Ongoing; Reported quantity: 1 unit; Lot: UDI-DI: N/A Serial Numbers: 43006. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips North America Llc
Distributor
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Lot
UDI-DI: N/A Serial Numbers: 43006
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Achieva 3.0T for PET -Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): 781477 recalled?

Yes. FDA published recall Z-1840-2024 on May 29, 2024. Reported hazard: The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

What should I do if I have this product?

Voluntary: Firm initiated · Letter