FDAFDA · Food and drugs·Recall number Z-1847-2024

Ingenia 1.5T S Model Number (REF): 781347

Published May 29, 2024 · Recall initiated: May 3, 2024

Summary

FDA records medical device Class II recall Z-1847-2024, published May 29, 2024 and initiated May 3, 2024. Recalling firm: Philips North America Llc; Status: Ongoing; Reported quantity: 139 units; Lot: UDI-DI: (01)00884838068421(21) Serial Numbers: 81003 81050 81051 81090 81111 81010 81027 81028 81030 81035 81063 81106 81093 81105 81116 81007 81004 81006 81009 81011 81016 81017 81018 81019 81020 81023 81024 81031 81037 81041 81055 81056 81057 81060 81062 81064 81070 81074 81078 81112 81115 81122 81146 81147 81153 81156 81158 81013 81044 81040 81047 81059 81167 81170 81171 81008 81108 81148 81133 81164 81001 81086 81079 81118 81138 81139 81140 81151 81175 81184 81069 81061 81065 81098 81085 81091 81092 81032 81043 81129 81131 81149 81155 81162 81165 81172 81180 81084 81087 81088 81094 81095 81100 81127 81132 81174 81157 81176 81071 81052 81067 81039 81038 81179 81185 81117 81173 81154 81025 81097 81152 81015 81029 81033 81034 81042 81046 81048 81049 81058 81072 81080 81103 81113 81123 81128 81137 81045 81075 81076 81077 81096 81114 81124 81126 81145 81159 81178. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips North America Llc
Distributor
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Lot
UDI-DI: (01)00884838068421(21) Serial Numbers: 81003 81050 81051 81090 81111 81010 81027 81028 81030 81035 81063 81106 81093 81105 81116 81007 81004 81006 81009 81011 81016 81017 81018 81019 81020 81023 81024 81031 81037 81041 81055 81056 81057 81060 81062 81064 81070 81074 81078 81112 81115 81122 81146 81147 81153 81156 81158 81013 81044 81040 81047 81059 81167 81170 81171 81008 81108 81148 81133 81164 81001 81086 81079 81118 81138 81139 81140 81151 81175 81184 81069 81061 81065 81098 81085 81091 81092 81032 81043 81129 81131 81149 81155 81162 81165 81172 81180 81084 81087 81088 81094 81095 81100 81127 81132 81174 81157 81176 81071 81052 81067 81039 81038 81179 81185 81117 81173 81154 81025 81097 81152 81015 81029 81033 81034 81042 81046 81048 81049 81058 81072 81080 81103 81113 81123 81128 81137 81045 81075 81076 81077 81096 81114 81124 81126 81145 81159 81178
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Ingenia 1.5T S Model Number (REF): 781347 recalled?

Yes. FDA published recall Z-1847-2024 on May 29, 2024. Reported hazard: The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

What should I do if I have this product?

Voluntary: Firm initiated · Letter