NobelParallel Conical Connection NP, 3.75x10mm (Endosseous Dental Implant); REF 37965; UDI (01)07332747091441(10)12129104(17)240203 - Product Usage: are intended to be used in the upper or lower jaw bone (osseointegration) and used for anchoring or supporting tooth replacements to restore chewing function.

Published May 13, 2020 · Recall initiated: August 28, 2019

Summary

FDA records medical device Class II recall Z-1851-2020, published May 13, 2020 and initiated August 28, 2019. Recalling firm: Nobel Biocare Usa Llc; Status: Terminated; Reported quantity: 228 pieces; Lot: Lot:12129104. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The top label of article no. 37965, lot 12129104 (immediate packaging of endosseous dental implants ) may incorrectly indicate an implant length of 7mm, which is shorter than the actual length of the implant (10mm).

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Nobel Biocare Usa Llc
Distributor
International distribution - the country Russia only.
Lot
Lot:12129104
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was NobelParallel Conical Connection NP, 3.75x10mm (Endosseous Dental Implant); REF 37965; UDI (01)07332747091441(10)12129104(17)240203 - Product Usage: are intended to be used in the upper or lower jaw bone (osseointegration) and used for anchoring or supporting tooth replacements to restore chewing function. recalled?

Yes. FDA published recall Z-1851-2020 on May 13, 2020. Reported hazard: The top label of article no. 37965, lot 12129104 (immediate packaging of endosseous dental implants ) may incorrectly indicate an implant length of 7mm, which is shorter than the actual length of the implant (10mm).

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail