Go Bed II+ Model FL28C Beds are intended for medical purposes in both general patient care as well as critical care environments. Each uses built-in electric motors and controls which can be operated by the patient, or caregiver, to adjust the height and/or surface contour of the bed.The devices listed above also include movable and latchable side rails which can be lowered or raised to help prevent patient fall and subsequent injury. The Beds include an optional Scale (Models: FL20E, FL25E, FL28C, FL28EX) as well as an optional Bed Exit (1 Zone) or Chaperone
Published July 4, 2012 · Recall initiated: May 17, 2012
Summary
FDA records medical device Class II recall Z-1860-2012, published July 4, 2012 and initiated May 17, 2012. Recalling firm: Stryker Medical Division of Stryker Corporation; Status: Terminated; Reported quantity: 8009; Lot: Model FL28C: Serial Number R25210, R21286, R20899, R16969, R16968, R15495, R15494, R15493, R15492, R15491, R15490, R15489, R15488, R15487, R15486, R15485, R15484, R15483, R15482, R15481, R15480, R15479, R15478, R15447, R15446, R15445, R15444, R15443, R15442, R15441, R15440, R15439, R15438, R15437, R15436, R15435, R15434, R15433, R15432, R15431, R15430, R15429, R15428, R15427, R15426, R15425, R15424, R15423, R15422, R15421, R15420, R15419, R15418, R15417, R15132, R15131, R15130, R15129, R15128, R15127, R15126, R15125, R15124, R15123, R15122, R15121, R15120, R15119, R15118, R15117, R15116, R15115, R15114, R15113, R15112, R15111, R15110, R15109, R15108, R15107, R15106, R15105, R15104, R15103, R15102, R15028, R15027, R15026, R15025, R15024, R15023, R15022, R15021, R15020, R15019, R15018, R15017, R15016, R15015, R15014, R15013, R15012, R15011, R15010, R15009, R15008, R15007, R15006, R15005, R15004, R15003, R15002, R15001, R15000, R14999, R14998, R14956, R14955, R14954, R14953, R14952, R14951, R14950, R14949, R14948, R14947, R14946, R14945, R14944, R14943, R14942, R14941, R14940, R14939, R14938, R14937, R14936, R14935, R14934, R14933, R14932, R14931, R14930, R14929, R14928, R14927, R14926, R14925, R14924, R14923, R14922, R14921, R14920, R14919, R14918, R14917, R14916, R14915, R14914, R14913, R14912, R14911, R14889, R14888, R14887, R14886, R14885, R14884, R14883, R14882, R14881, R14880, R14879, R14878, R14877, R14876, R14875, R14874, R14873, R14872, R14871, R14870, R14869, R14868, R14867, R14866, R14865, R14864, R14863, R14862, R14861, R14860, R14859, R14858, R14857, R14856, R14855, R14854, R14853, R14846, R14845, R14844, R14843, R14842, R14841, R14840, R14839, R14838, R14837, R14836, R14835, R14834, R14833, R14832, R14831, R14830, R14829, R14828, R14827, R14826, R14825, R14824, R14823, R14822, R14821, R14820, R14819, R14818, R14817, R14816, R14815, R14814, R14813, R14812, R14811, R14810, R14809, R14808, R14807, R14806, R14805, R14804, R14803, R14802, R14801, R14800, R14799, R14798, R14737, R14736, R14735, R14734, R14733, R14732, R14731, R14730, R14729, R14728, R14727, R14726, R14725, R14724, R14723, R14722, R14721, R14720, R14719, R14718, R14717, R14716, R14715, R14714, R14713, R14712, R14711, R14710, R14709, R14708, R14707, R14706, R14705, R14704, R14703, R14702, R14701, R14700, R14699, R14698, R14697, R14696, R14695, R14694, R14693, R14692, R14691, R14690, R14689, R14688, R14687, R14686, R14685, R14684, R14683, R14682, R14681, R14680, R14679, R14678, R14677, R14676, R14675, R14674, R14673, R14672, R14671, R14670, R14669, R14668, R14667, R14666, R14665, R14664, R14663, R14662, R14661, R14660, R14659, R14658, R14657, R14656, R14655, R14654, R14653, R14652, R14651, R14650, R14649, R14648, R14647, R14646, R14645, R14644, R14643, R14642, R14641, R14640, R14639, R14638, R14637, R14636, R14635, R14634, R14633, R14632, R14631, R14630, R14629, R14628, R14627, R14626, R14625, R14624, R14623, R14622, R14621, R14620, R14619, R14618, R14617, R14616, R14615, R14614, R14613, R14604, R14603, R14602, R14601, R14600, R14599, R14598, R14597, R14596, R14595, R14594, R14593, R14592, R14591, R14590, R14589, R14588, R14587, R14586, R14585, R14584, R14583, R14582, R14581, R14580, R14579, R14578, R14577, R14576, R14575, R14574, R14573, R14572, R14571, R14570, R14569, R14568, R14498, R14497, R14496, R14495, R14494, R14493, R14492, R14491, R14490, R14489, R14488, R14487, R14486, R14485, R14484, R14483, R14482, R14481, R14480, R14479, R14478, R14477, R14476, R14475, R14474, R14473, R14472, R14471, R14470, R14469, R14468, R14467, R14466, R14465, R14464, R14463, R14462, R14461, R14460, R14459, R14458, R14457, R14456, R14455, R14454, R14453, R14452, R14451, R14450, R14449, R14448, R14447, R14439, R14438, R14437, R14436, R14435, R14434, R14433, R14432, R14431, R14430, R14429, R14428, R14427, R14426, R14425, R14424, R14395, R14394, R14393, R14392, R14391, R14390, R14389, R14388, R14387, R14375, R14374, R14373, R14372, R14371, R14370, R14369, R14368, R14367, R143. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
An adverse trend of scale and bed exit service reports/complaints were detected for the GoBed+, GoBed II, Rose, and MA204 model beds. Investigation found an abnormally high failure rate of loadcells, an integral component in scale and bed exit function. The scale faults are manifested as a fluctuating scale reading, a scale error causing the scale system to be disabled until serviced, or an inac
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Medical Division of Stryker Corporation
- Distributor
- Worldwide distribution:USA (nationwide) and countries including: Argentina, Bertec, Brasil, Brazil, Canada, Chile, China, India, Latin America, Mexico,Osteonics SA, Polska, Portugal, Puerto Rico, and Singapore.
- Lot
- Model FL28C: Serial Number R25210, R21286, R20899, R16969, R16968, R15495, R15494, R15493, R15492, R15491, R15490, R15489, R15488, R15487, R15486, R15485, R15484, R15483, R15482, R15481, R15480, R15479, R15478, R15447, R15446, R15445, R15444, R15443, R15442, R15441, R15440, R15439, R15438, R15437, R15436, R15435, R15434, R15433, R15432, R15431, R15430, R15429, R15428, R15427, R15426, R15425, R15424, R15423, R15422, R15421, R15420, R15419, R15418, R15417, R15132, R15131, R15130, R15129, R15128, R15127, R15126, R15125, R15124, R15123, R15122, R15121, R15120, R15119, R15118, R15117, R15116, R15115, R15114, R15113, R15112, R15111, R15110, R15109, R15108, R15107, R15106, R15105, R15104, R15103, R15102, R15028, R15027, R15026, R15025, R15024, R15023, R15022, R15021, R15020, R15019, R15018, R15017, R15016, R15015, R15014, R15013, R15012, R15011, R15010, R15009, R15008, R15007, R15006, R15005, R15004, R15003, R15002, R15001, R15000, R14999, R14998, R14956, R14955, R14954, R14953, R14952, R14951, R14950, R14949, R14948, R14947, R14946, R14945, R14944, R14943, R14942, R14941, R14940, R14939, R14938, R14937, R14936, R14935, R14934, R14933, R14932, R14931, R14930, R14929, R14928, R14927, R14926, R14925, R14924, R14923, R14922, R14921, R14920, R14919, R14918, R14917, R14916, R14915, R14914, R14913, R14912, R14911, R14889, R14888, R14887, R14886, R14885, R14884, R14883, R14882, R14881, R14880, R14879, R14878, R14877, R14876, R14875, R14874, R14873, R14872, R14871, R14870, R14869, R14868, R14867, R14866, R14865, R14864, R14863, R14862, R14861, R14860, R14859, R14858, R14857, R14856, R14855, R14854, R14853, R14846, R14845, R14844, R14843, R14842, R14841, R14840, R14839, R14838, R14837, R14836, R14835, R14834, R14833, R14832, R14831, R14830, R14829, R14828, R14827, R14826, R14825, R14824, R14823, R14822, R14821, R14820, R14819, R14818, R14817, R14816, R14815, R14814, R14813, R14812, R14811, R14810, R14809, R14808, R14807, R14806, R14805, R14804, R14803, R14802, R14801, R14800, R14799, R14798, R14737, R14736, R14735, R14734, R14733, R14732, R14731, R14730, R14729, R14728, R14727, R14726, R14725, R14724, R14723, R14722, R14721, R14720, R14719, R14718, R14717, R14716, R14715, R14714, R14713, R14712, R14711, R14710, R14709, R14708, R14707, R14706, R14705, R14704, R14703, R14702, R14701, R14700, R14699, R14698, R14697, R14696, R14695, R14694, R14693, R14692, R14691, R14690, R14689, R14688, R14687, R14686, R14685, R14684, R14683, R14682, R14681, R14680, R14679, R14678, R14677, R14676, R14675, R14674, R14673, R14672, R14671, R14670, R14669, R14668, R14667, R14666, R14665, R14664, R14663, R14662, R14661, R14660, R14659, R14658, R14657, R14656, R14655, R14654, R14653, R14652, R14651, R14650, R14649, R14648, R14647, R14646, R14645, R14644, R14643, R14642, R14641, R14640, R14639, R14638, R14637, R14636, R14635, R14634, R14633, R14632, R14631, R14630, R14629, R14628, R14627, R14626, R14625, R14624, R14623, R14622, R14621, R14620, R14619, R14618, R14617, R14616, R14615, R14614, R14613, R14604, R14603, R14602, R14601, R14600, R14599, R14598, R14597, R14596, R14595, R14594, R14593, R14592, R14591, R14590, R14589, R14588, R14587, R14586, R14585, R14584, R14583, R14582, R14581, R14580, R14579, R14578, R14577, R14576, R14575, R14574, R14573, R14572, R14571, R14570, R14569, R14568, R14498, R14497, R14496, R14495, R14494, R14493, R14492, R14491, R14490, R14489, R14488, R14487, R14486, R14485, R14484, R14483, R14482, R14481, R14480, R14479, R14478, R14477, R14476, R14475, R14474, R14473, R14472, R14471, R14470, R14469, R14468, R14467, R14466, R14465, R14464, R14463, R14462, R14461, R14460, R14459, R14458, R14457, R14456, R14455, R14454, R14453, R14452, R14451, R14450, R14449, R14448, R14447, R14439, R14438, R14437, R14436, R14435, R14434, R14433, R14432, R14431, R14430, R14429, R14428, R14427, R14426, R14425, R14424, R14395, R14394, R14393, R14392, R14391, R14390, R14389, R14388, R14387, R14375, R14374, R14373, R14372, R14371, R14370, R14369, R14368, R14367, R143
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAProCuity bed series, model number 3009, item number: 3009PX-LE-400.December 13, 2023
- FDAProform- support surface is used to assist in the prevention and treatment of all categories/stages ofAugust 18, 2021
- FDAPrime Fifth Wheel Stretcher-stretcher may be used for minor procedures and short-term stay (treatment andAugust 18, 2021
- FDAThe Spirit Plus Bed, Model 5600 Spirit Plus /with high rail siderail optionNovember 21, 2018
- FDAGoBed+ . Model FL20E Beds are intended for medical purposes in both general patient care as well as criticalJuly 4, 2012
- FDAStryker Rose Bed Model (FL14E) Beds are intended for medical purposes in both general patient care as well asJuly 4, 2012
Questions about this recall
Was Go Bed II+ Model FL28C Beds are intended for medical purposes in both general patient care as well as critical care environments. Each uses built-in electric motors and controls which can be operated by the patient, or caregiver, to adjust the height and/or surface contour of the bed.The devices listed above also include movable and latchable side rails which can be lowered or raised to help prevent patient fall and subsequent injury. The Beds include an optional Scale (Models: FL20E, FL25E, FL28C, FL28EX) as well as an optional Bed Exit (1 Zone) or Chaperone recalled?
Yes. FDA published recall Z-1860-2012 on July 4, 2012. Reported hazard: An adverse trend of scale and bed exit service reports/complaints were detected for the GoBed+, GoBed II, Rose, and MA204 model beds. Investigation found an abnormally high failure rate of loadcells, an integral component in scale and bed exit function. The scale faults are manifested as a fluctuating scale reading, a scale error causing the scale system to be disabled until serviced, or an inac
What should I do if I have this product?
Voluntary: Firm initiated · Letter