FDAFDA · Food and drugs·Recall number Z-1861-2024
SmartPath to dStream for 3.0T Model Number (REF): 782145
Published May 29, 2024 · Recall initiated: May 3, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Philips North America Llc
- Distributor
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
- Lot
- UDI-DI: N/A Serial Numbers: 17377 34114 38294
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States