LAP LASER PICTOR 3D Multi Color Laser System PICTOR 3D is a LAP laser system which uses multi projection format that defines isocenters, field and MLC contours along with a 3D patient view.
Published July 2, 2014 · Recall initiated: March 21, 2014
Summary
FDA records medical device Class II recall Z-1866-2014, published July 2, 2014 and initiated March 21, 2014. Recalling firm: LAP of America Laser Applications, LLC; Status: Terminated; Reported quantity: 1 system; Lot: Product Code: 0007569-0001. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
While projecting yellow lines it was notice that the yellow projection has the potential to splits into two separate red and green projections. In this case the difference between the red and green projection was up to 2.5mm.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- LAP of America Laser Applications, LLC
- Distributor
- North Carolina
- Lot
- Product Code: 0007569-0001
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIngenia 1.5T, magnetic resonance imaging system.July 2, 2014
- FDASiemens StreamLAB(R) Analytical Workcell; Catalog number 746002.901. Designed to automate sample handling andJuly 2, 2014
- FDAIngenia 3.0T, magnetic resonance imaging system.July 2, 2014
- FDAAccell Evo3C Demineralized Bone Matrix, 5 cc, Model Number: 02-6000-050 , Lot Number: 132160. Product Usage:July 2, 2014
- FDAIngenia 1.5T R5, magnetic resonance imaging system.July 2, 2014
- FDACARINAsim MAN-1236 Rev 2 and CARINAiso MAN-1237 Rev 2 laser software. LAP isocenter marking control software.June 11, 2014
Questions about this recall
Was LAP LASER PICTOR 3D Multi Color Laser System PICTOR 3D is a LAP laser system which uses multi projection format that defines isocenters, field and MLC contours along with a 3D patient view. recalled?
Yes. FDA published recall Z-1866-2014 on July 2, 2014. Reported hazard: While projecting yellow lines it was notice that the yellow projection has the potential to splits into two separate red and green projections. In this case the difference between the red and green projection was up to 2.5mm.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit