Back to search
FDAFDA · Food and drugs·Recall number Z-1867-2012

BCS(R) Behring Coagulation System, Device catalog No. OVIO03 Product Usage: Multipurpose system for In Vitro coagulation studies

Published July 4, 2012 · Recall initiated: December 1, 2010

Summary

FDA records medical device Class II recall Z-1867-2012, published July 4, 2012 and initiated December 1, 2010. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Terminated; Reported quantity: 186; Lot: Device catalog No. OVIO03 (not lot specific). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Firm has confirmed that if the Cleaner SCS vial becomes empty after processing a test for Antithrombin assay using the INNOVANCE(R) Antithrombin kit, and the test immediately following is an APTT based clotting assay several results may be shortened.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Healthcare Diagnostics, Inc.
Distributor
Worldwide Distribution - US (nationwide) including the states of: AZ, AR, CA, CO, CT, FL, GA, ID, IL, IN, IO, KS, KY, LA, MD, MA, MI, MS, MN, MT, NE, NH, NJ, NY, NC, ND, OH, OK, OR, PA, TN, TX, UT, VA, WA, WV, and WI and the countries of: Australia, France, Canary Islands, Spain, Belgium, Germany, Switzerland, Netherlands, Italy, Guadeloupe, Ceuta, Austria, Slovenia, Croatia, Saudi Arabia, Turkey, Greece, and San Marino.
Lot
Device catalog No. OVIO03 (not lot specific)
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was BCS(R) Behring Coagulation System, Device catalog No. OVIO03 Product Usage: Multipurpose system for In Vitro coagulation studies recalled?

Yes. FDA published recall Z-1867-2012 on July 4, 2012. Reported hazard: Firm has confirmed that if the Cleaner SCS vial becomes empty after processing a test for Antithrombin assay using the INNOVANCE(R) Antithrombin kit, and the test immediately following is an APTT based clotting assay several results may be shortened.

What should I do if I have this product?

Voluntary: Firm initiated · Letter