FDAFDA · Food and drugs·Recall number Z-1870-2021

Atellica CH 930 Analyzer with Photometer Lamp (SMN 11075676) Siemens Material Number: 11067000

Published June 23, 2021 · Recall initiated: May 4, 2021

Summary

FDA records medical device Class II recall Z-1870-2021, published June 23, 2021 and initiated May 4, 2021. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Ongoing; Reported quantity: 2,153 units; Lot: Software version: 1.24.0 SP1 or lower. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Photometer Lamp May Reach Saturation Without Flagging Results, may generate erroneously elevated or depressed patient results

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Healthcare Diagnostics, Inc.
Distributor
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Burkina Faso, Canada, Chile, Colombia, Croatia Cura®ao, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, P.R. China Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., Uganda, United Kingdom, Uruguay, Vatikancity, Vietnam.
Lot
Software version: 1.24.0 SP1 or lower
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Atellica CH 930 Analyzer with Photometer Lamp (SMN 11075676) Siemens Material Number: 11067000 recalled?

Yes. FDA published recall Z-1870-2021 on June 23, 2021. Reported hazard: Photometer Lamp May Reach Saturation Without Flagging Results, may generate erroneously elevated or depressed patient results

What should I do if I have this product?

Voluntary: Firm initiated · Letter