Atellica CH 930 Analyzer with Photometer Lamp (SMN 11075676) Siemens Material Number: 11067000
Published June 23, 2021 · Recall initiated: May 4, 2021
Summary
FDA records medical device Class II recall Z-1870-2021, published June 23, 2021 and initiated May 4, 2021. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Ongoing; Reported quantity: 2,153 units; Lot: Software version: 1.24.0 SP1 or lower. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Photometer Lamp May Reach Saturation Without Flagging Results, may generate erroneously elevated or depressed patient results
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Burkina Faso, Canada, Chile, Colombia, Croatia Cura®ao, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, P.R. China Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., Uganda, United Kingdom, Uruguay, Vatikancity, Vietnam.
- Lot
- Software version: 1.24.0 SP1 or lower
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAtellica CH Revised C-Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223;October 7, 2026
- FDABeckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mLAugust 26, 2026
- FDASyva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mLAugust 26, 2026
- FDASyva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219ULAugust 12, 2026
- FDAPremierZEN Gold 7000 capsule, 1-count blister card, packaged in 12 cards per box (UPC 7 28175 42183 2),June 30, 2021
- FDAAdult-Pediatric Electrostatic Filter HME, Item Code 352/5877Z - Product Usage: indicated for single use onJune 23, 2021
Questions about this recall
Was Atellica CH 930 Analyzer with Photometer Lamp (SMN 11075676) Siemens Material Number: 11067000 recalled?
Yes. FDA published recall Z-1870-2021 on June 23, 2021. Reported hazard: Photometer Lamp May Reach Saturation Without Flagging Results, may generate erroneously elevated or depressed patient results
What should I do if I have this product?
Voluntary: Firm initiated · Letter