Shaver Handpiece Sterilization Tray. Model number 272-700-000. For sterilization of Stryker Endoscopy arthroscopic shaver.
Published July 2, 2014 · Recall initiated: May 19, 2014
Summary
FDA records medical device Class II recall Z-1872-2014, published July 2, 2014 and initiated May 19, 2014. Recalling firm: Stryker Endoscopy; Status: Terminated; Reported quantity: 1992 trays; Lot: Part Number 272-700-000: Lot number: 20476.1 23541 23837 26403 26404 26405 26676 27009 27518 28341 30248 34355 35932 36756 39705 05J034834 06con05774 06CON06360 06CON06846 06con06869 06con06960 06con08587 25067-1-12 26882-1-1 26882-1-11 26882-1-12 26882-1-13 26882-1-14 35200-1-1 36610-1-1 38820-1-1 39078-1-1 40504-2-1 40675-1-1 41231-1-1 41827-1-1 42085-1-1 42375-1-1 42527-1-1 42789-1-1 43103-1-1 43589-1-1 43716-1-1 43716-1-3 43716-1-4 44568-1-10 44568-1-11 44568-1-2 44568-1-3 44568-1-4 44568-1-6 44568-1-7 44568-1-8 44568-1-9 46227-1-1 47731-1-1 48595-1-1 48596-1-1 48597-1-1 48598-1-1 48599-1-1 48600-1-1 50101-1-1 50628-1-1 50828-1-1 51237-1-1 52439-1-1 55035-1-1 55422-1-1 55462-1-1 56456-1-1 R471611 SS15329. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
STERRAD 100S parameters provided in the Shaver Handpiece reprocessing guide (P10299 Revision A) did not fully consider all worst-case scenarios during validation testing.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Endoscopy
- Distributor
- Worldwide Distribution.
- Lot
- Part Number 272-700-000: Lot number: 20476.1 23541 23837 26403 26404 26405 26676 27009 27518 28341 30248 34355 35932 36756 39705 05J034834 06con05774 06CON06360 06CON06846 06con06869 06con06960 06con08587 25067-1-12 26882-1-1 26882-1-11 26882-1-12 26882-1-13 26882-1-14 35200-1-1 36610-1-1 38820-1-1 39078-1-1 40504-2-1 40675-1-1 41231-1-1 41827-1-1 42085-1-1 42375-1-1 42527-1-1 42789-1-1 43103-1-1 43589-1-1 43716-1-1 43716-1-3 43716-1-4 44568-1-10 44568-1-11 44568-1-2 44568-1-3 44568-1-4 44568-1-6 44568-1-7 44568-1-8 44568-1-9 46227-1-1 47731-1-1 48595-1-1 48596-1-1 48597-1-1 48598-1-1 48599-1-1 48600-1-1 50101-1-1 50628-1-1 50828-1-1 51237-1-1 52439-1-1 55035-1-1 55422-1-1 55462-1-1 56456-1-1 R471611 SS15329.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080123 Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, ENDO METZENBAUM SCISSORS, CURVED, P/N 0250080267. Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, ALLIGATOR FORCEPS, P/N 0250080316. Laparoscopic Manual Instruments are intended for cutting, holding,December 24, 2014
- FDAPKG, 5MM X 33CM, INSERT, FENESTRATED GRASPER, P/N 0250080755 Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAIS4000 da Vinci Xi - Endoscope accessory; 0 Degree, 8mm Endoscope. Intuitive Surgical Inc.July 2, 2014
- FDAGlucose Hexokinase Liquid Reagent For the in vitro quantitative determination of Glucose in serum.July 2, 2014
Questions about this recall
Was Shaver Handpiece Sterilization Tray. Model number 272-700-000. For sterilization of Stryker Endoscopy arthroscopic shaver. recalled?
Yes. FDA published recall Z-1872-2014 on July 2, 2014. Reported hazard: STERRAD 100S parameters provided in the Shaver Handpiece reprocessing guide (P10299 Revision A) did not fully consider all worst-case scenarios during validation testing.
What should I do if I have this product?
Voluntary: Firm initiated · Letter