Sonopet Ultrasonic Aspirator Console (Console 110V with Pedal & Pole) Product Usage: The Sonopet Console is indicated for use in surgical procedures where fragmentation, emulsification, and aspiration of soft and hard tissue is desirable, including neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, plastic and reconstructive surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, and thoracoscopic surgery.
Published August 14, 2013 · Recall initiated: June 24, 2013
Summary
FDA records medical device Class II recall Z-1878-2013, published August 14, 2013 and initiated June 24, 2013. Recalling firm: Stryker Instruments Div. of Stryker Corporation; Status: Terminated; Reported quantity: 215 units total; Lot: 5450-850-000 (Console 110V with Pedal & Pole) S/N 1131300134, 1131900074, 1200500134, 1203200244, 1210100114, 1131300144, 1131900084, 1200500144, 1203200254, 1210100124, 1131300154, 1131900094, 1200500154, 1203200274, 1210100134, 1131300164, 1131900104, 1200500164, 1203700954, 1210100144, 1131300174, 1131900114, 1200500174, 1203700964, 1210100154, 1131300184, 1131900124, 1200500184, 1203700974, 1210700014, 1131300194, 1132800114, 1200500194, 1203900024, 1210700024, 1131300204, 1132800124, 1200500204, 1204000044, 1210700034, 1131300214, 1132800134, 1201300494, 1204000054, 1210700044, 1131300224, 1132800144, 1201300544, 1207400154, 1210700054, 1131300234, 1132800154, 1201300754, 1207400164, 1213800114, 1131300244, 1132800164, 1201300764, 1207400174, 1213800124, 1131900014, 1132800174, 1201300774, 1207400184, 1213800134, 1131900024, 1132800184, 1201300784, 1208300014, 1213800144, 1131900034, 1132800194, 1201300794, 1208300024, 1213800154, 1131900044, 1132800204, 1203200214, 1208300034, 1131900054, 1200500114, 1203200224, 1208300044, 1131900064, 1200500124, 1203200234, 1208300054. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A supplied component exhibiting corrosion was built into the power supply boards of the Sonopet Consoles listed above. The issue is limited to 2 lots of the supplied component. The presence of the corrosion on the relay of the power supply board can potentially lead to a conduction defect which could cause the irrigation pump to stop working. The purpose of the irrigation pump is to supply sali
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Instruments Div. of Stryker Corporation
- Distributor
- Worldwide - USA Nationwide including the states of CA, CO, FL, HI, IA, IL, KS, KY, LA, MD, MI, MN, MS, NC, NH, NJ, NV, NY, OH, PA, TX, VA, ITALY, and the countries of AUSTRALIA, CANADA, UNITED KINGDOM, NETHERLANDS, INDIA, JAPAN, NEW ZEALAND, SWITZERLAND, HONG KONG, and TAIWAN.
- Lot
- 5450-850-000 (Console 110V with Pedal & Pole) S/N 1131300134, 1131900074, 1200500134, 1203200244, 1210100114, 1131300144, 1131900084, 1200500144, 1203200254, 1210100124, 1131300154, 1131900094, 1200500154, 1203200274, 1210100134, 1131300164, 1131900104, 1200500164, 1203700954, 1210100144, 1131300174, 1131900114, 1200500174, 1203700964, 1210100154, 1131300184, 1131900124, 1200500184, 1203700974, 1210700014, 1131300194, 1132800114, 1200500194, 1203900024, 1210700024, 1131300204, 1132800124, 1200500204, 1204000044, 1210700034, 1131300214, 1132800134, 1201300494, 1204000054, 1210700044, 1131300224, 1132800144, 1201300544, 1207400154, 1210700054, 1131300234, 1132800154, 1201300754, 1207400164, 1213800114, 1131300244, 1132800164, 1201300764, 1207400174, 1213800124, 1131900014, 1132800174, 1201300774, 1207400184, 1213800134, 1131900024, 1132800184, 1201300784, 1208300014, 1213800144, 1131900034, 1132800194, 1201300794, 1208300024, 1213800154, 1131900044, 1132800204, 1203200214, 1208300034, 1131900054, 1200500114, 1203200224, 1208300044, 1131900064, 1200500124, 1203200234, 1208300054.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASurgiCount+ System Application, Software Version 2.0.17, Catalog Number 0694-002-090. Used for counting,July 27, 2022
- FDANeptune E-SEP 165mm Blade Electrode, Catalog Number 0703-165-001 - Product Usage: The Neptune E-SEP is anMay 6, 2020
- FDAColor Cuff Disposable Tourniquet Cuff, 18 inch X 3 inch, 1BLA (Single Bladder), 1PRT(Single Port) QuickDecember 11, 2019
- FDAC2 Nerve Monitor, 8 Channel, 5140-508-280, UDI 07613327277319 Product Usage: The C2 Nerve Monitor System isOctober 9, 2019
- FDAMedtronic Mullins Transseptal Adult 7F, REF 008551. Intended for the introduction of cardiovascular cathetersAugust 14, 2013
- FDASonopet Ultrasonic Aspirator Console 5450-851-000(Console 100V) Product Usage: The Sonopet Console isAugust 14, 2013
Questions about this recall
Was Sonopet Ultrasonic Aspirator Console (Console 110V with Pedal & Pole) Product Usage: The Sonopet Console is indicated for use in surgical procedures where fragmentation, emulsification, and aspiration of soft and hard tissue is desirable, including neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, plastic and reconstructive surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, and thoracoscopic surgery. recalled?
Yes. FDA published recall Z-1878-2013 on August 14, 2013. Reported hazard: A supplied component exhibiting corrosion was built into the power supply boards of the Sonopet Consoles listed above. The issue is limited to 2 lots of the supplied component. The presence of the corrosion on the relay of the power supply board can potentially lead to a conduction defect which could cause the irrigation pump to stop working. The purpose of the irrigation pump is to supply sali
What should I do if I have this product?
Voluntary: Firm initiated · Letter