Back to search
FDAFDA · Food and drugs·Recall number Z-1884-2024

Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24

Published May 29, 2024 · Recall initiated: April 11, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
FUJIFILM Healthcare Americas Corporation
Distributor
US Nationwide distribution.
Lot
UDI-DI: (01) 04580292771704 Noblus Ultrasound Imaging System AC Adapter Serial Number Before K17160001 (Excluding K17160001)
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card