Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24
Published May 29, 2024 · Recall initiated: April 11, 2024
Summary
FDA records medical device Class II recall Z-1884-2024, published May 29, 2024 and initiated April 11, 2024. Recalling firm: FUJIFILM Healthcare Americas Corporation; Status: Ongoing; Reported quantity: 418 units; Lot: UDI-DI: (01) 04580292771704 Noblus Ultrasound Imaging System AC Adapter Serial Number Before K17160001 (Excluding K17160001). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- FUJIFILM Healthcare Americas Corporation
- Distributor
- US Nationwide distribution.
- Lot
- UDI-DI: (01) 04580292771704 Noblus Ultrasound Imaging System AC Adapter Serial Number Before K17160001 (Excluding K17160001)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFUJIFILM Synapse PACS software v7.4.310; radiological image processing systemSeptember 2, 2026
- FDAThe device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2October 2, 2024
- FDAFDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinalJune 5, 2024
- FDAStryker Blueprint Software, Catalog #BPUE001.May 29, 2024
- FDA8000-021-002, Scopis® ENT Software with TGS® with Sofware Version: NOVA 3.6.0 RC16. Scopis is aMay 29, 2024
- FDASynapse PACS - Version 7.2.000March 13, 2024
Questions about this recall
Was Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24 recalled?
Yes. FDA published recall Z-1884-2024 on May 29, 2024. Reported hazard: AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).
What should I do if I have this product?
Voluntary: Firm initiated · Letter