Gas Delivery Engine (GDE), 16650A A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined percentage of oxygen in the breathing gas.
Published June 22, 2016 · Recall initiated: May 17, 2016
Summary
FDA records medical device Class I recall Z-1896-2016, published June 22, 2016 and initiated May 17, 2016. Recalling firm: Carefusion 211 Inc dba Carefusion; Status: Terminated; Reported quantity: 922 units total (501 units in US). Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Carefusion 211 Inc dba Carefusion
- Distributor
- Worldwide Distribution - US (Nationwide) and the countries of Australia, Hungary, Paraguay, Belgium, Indonesia, Philippines, Bolivia, Ireland, Poland, Bermuda, India, Saudi Arabia, Canada, Japan, Singapore, Chile, Jordon, South Africa, China, Kuwait, Taiwan, Ecuador, Malaysia, Thailand, Egypt, Mexico, Turkey, France, Netherlands, United Kingdom, Guatemala, and Nicaragua.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGas Delivery Engine (GDE)-refurbished, R16650A A continuous ventilator intended to mechanically control orJune 22, 2016
- FDAAVEA Standard with Compressor ventilator-refurbished, R17312-xx (xx = 0-14) A continuous ventilator intendedJune 22, 2016
- FDADucktrap Smoked Mussels in both 6oz and 16oz TubsJune 22, 2016
- FDABioSync Anatomic Cotton Wedge, 16 x 4.5 mm.; Used for internal bone fixation for bone fractures, fusions orJune 22, 2016
- FDARose Red Organic Ginger Steam Filter Bag & Organic Ground Ginger Steam packaged in a poly-lined corrugatedJune 22, 2016
- FDAProduct Name: AVEA® Ventilator. All Models. The affected devices are cleared under K073069 pm 1/25/2008 andOctober 9, 2013
Questions about this recall
Was Gas Delivery Engine (GDE), 16650A A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. recalled?
Yes. FDA published recall Z-1896-2016 on June 22, 2016. Reported hazard: CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
What should I do if I have this product?
Voluntary: Firm initiated · Letter