Philips Allura Xper FD20 Biplane system with an ADN7NT patient table. Model Number: 722013, Software Version Number 8.1.100.

Published June 5, 2024 · Recall initiated: May 16, 2024

Summary

FDA records medical device Class II recall Z-1897-2024, published June 5, 2024 and initiated May 16, 2024. Recalling firm: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.; Status: Ongoing; Reported quantity: 1 unit (OUS); Lot: Product number: 722013, UDI/DI: 00884838059061; Serial Number: 232. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Distributor
Domestic: OK International: India, Spain, U.A.E.
Lot
Product number: 722013, UDI/DI: 00884838059061; Serial Number: 232.
Risk classification
Class II
Area (FDA)
device
Country
Netherlands