Adult Divided Cannula end tidal C)2 sampling w/male luer-lock connector. 7 ft, oxygen tube, 7 ft. CO2 tube RX only.For Single Patient Use only. Distributed by Tri-Anium Assembled in Mexico. Usage: Delivery of Anesthesia.
Published July 4, 2012 · Recall initiated: May 30, 2012
Summary
FDA records medical device Class II recall Z-1898-2012, published July 4, 2012 and initiated May 30, 2012. Recalling firm: SALTER LABS; Status: Terminated; Reported quantity: 242 cases; Lot: Salter Part Number SO-1296, Tri-Anium Part numaber: 301-P3600; Lot number: 032612. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A complaint was received, and later verified, that product labeled as SO-1296 with male luer lock connector actually contained product with a female luer lock connector.
What to do
Voluntary: Firm initiated · Telephone
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- SALTER LABS
- Distributor
- Nationwide distribution: USA including states of: FL, IN, KS, PA and TX.
- Lot
- Salter Part Number SO-1296, Tri-Anium Part numaber: 301-P3600; Lot number: 032612.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASalter Labs EtCO2 Sample line 15' (4.6 M) tube, male luer connectors. Rx only; Made in Mexico;November 27, 2013
- FDAEthicon Surgiflo Hemostatic Matrix Kit with Thrombin Plus Flex Tip Distributed by Ethicon Ethicon Inc.,July 4, 2012
- FDAMULTIGENT Acetaminophen for use with Architect/Aeroset Cat. No. 2K99-20. Assay is intended for theJuly 4, 2012
- FDAPhilips HeartStart MRx monitor/defibrillators Models: M3535A, M3536A, M3536M4, M3536M5, and M3536M6 ProductJuly 4, 2012
- FDAPTS Panels Creatinine test strips. PTS Panels Creatinine Test Strips are intended to measure creatinine inJuly 4, 2012
- FDABigliani/Flatow The Complete Shoulder Solution Trabecular Metal Glenoid Instrument Drill with Stop, 6 mm; theJuly 4, 2012
Questions about this recall
Was Adult Divided Cannula end tidal C)2 sampling w/male luer-lock connector. 7 ft, oxygen tube, 7 ft. CO2 tube RX only.For Single Patient Use only. Distributed by Tri-Anium Assembled in Mexico. Usage: Delivery of Anesthesia. recalled?
Yes. FDA published recall Z-1898-2012 on July 4, 2012. Reported hazard: A complaint was received, and later verified, that product labeled as SO-1296 with male luer lock connector actually contained product with a female luer lock connector.
What should I do if I have this product?
Voluntary: Firm initiated · Telephone