MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3

Published June 11, 2025 · Recall initiated: April 23, 2025

Summary

FDA records medical device Class II recall Z-1905-2025, published June 11, 2025 and initiated April 23, 2025. Recalling firm: VIRTUAL INCISION CORPORATION; Status: Completed; Reported quantity: 3 units; Lot: UDI-DI (01)00850038042028; Lot Numbers: 1020040211, 1020040212, 1020040213. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Completed — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the hand controllers and the movement of the Surgical Minibot arms.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
VIRTUAL INCISION CORPORATION
Distributor
US Distribution to States: CA, FL, NE
Lot
UDI-DI (01)00850038042028; Lot Numbers: 1020040211, 1020040212, 1020040213
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3 recalled?

Yes. FDA published recall Z-1905-2025 on June 11, 2025. Reported hazard: The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the hand controllers and the movement of the Surgical Minibot arms.

What should I do if I have this product?

Voluntary: Firm initiated · Letter