REF: Z2560, 100" 10 Drop Admin Set w/Hand Pump, 2 Clave", Rotating Luer, 1 Pressure Infusion (400psig) Ext, Purple Clamp, UDI: (01)(01)00887709026889

Published May 13, 2020 · Recall initiated: October 4, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop cannula could result in inaccurate delivery of the intended infusion volume (under-delivery). This could result in under-delivery of critical medication which may lead to serious adverse health consequences.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
ICU Medical, Inc.
Distributor
US:AK,AL,AR,AZ,CA,CO,CT,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,SD,TN,TX,UT,VA,WA,WI,WY OUS: Canada *** "REF" numbers AH7117 and B30211 were not marketed or distributed within the US***
Lot
Lot numbers: 4172061 and 4173149
Risk classification
Class II
Area (FDA)
device
Country
United States