Back to search
FDAFDA · Food and drugs·Recall number Z-1916-2024

STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screw implants during orthopaedic surgery. Part Number: 03.130.010

Published June 5, 2024 · Recall initiated: April 25, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Stardrive Screwdrivers manufactured without a 5-degree relief cut around the tip and will not engage with their respective 1.3mm and 1.5mm screws, may result in surgical delay

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Synthes (USA) Products LLC
Distributor
Worldwide distribution - US Nationwide and the country of Canada.
Lot
UDI-DI: 10886982075192 Lot Numbers: 97P6250, 97P6251, 97P6252
Risk classification
Class II
Area (FDA)
device
Country
United States