bioMerieux Vitek 2 Systems Software version 6.01 software, used with the VITEK 2 automated system. Kit, UPDT V2S 6.01 DA, SV. Part number: 414567. The product is used with the VITEK 2 automated system which consists of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast.
Published August 14, 2013 · Recall initiated: June 14, 2013
Summary
FDA records medical device Class II recall Z-1918-2013, published August 14, 2013 and initiated June 14, 2013. Recalling firm: Biomerieux Inc; Status: Terminated; Reported quantity: 12,384 total kits; Lot: version 6.01. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A product issue has been identified for customers using software version 6.01 and are interfaced with a Laboratory Information System (LIS). During the update of software from version 5.xx to 6.01, the Bid-directional Computer interface (BCI) antibiotic codes are not updated properly for Cefovecin, Prulifloxacin and Ceftaroline. For the referenced antibiotics, the firm code in the reference table
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Biomerieux Inc
- Distributor
- Worldwide distribution: US (nationwide) and countries of: Argentina, Australia, Austria, Bahrain, Belarus, Benelux, Brazil, Bulgaria, Canada, Chile China, Costa Rica, Croatia, Czech Republic, Dominican Republic, Dutch Antilles, Ecuador, Egypt, France, Germany, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Japan, Kazakhstan, Korea, Kuwait, Libya, Lithuania, Macedonia, Malaysia, Malta, Mexico, Morocco, Myanmar, Netherlands, Nicaragua, Nigeria, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, South Africa, Thailand, Tunisia, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.
- Lot
- version 6.01
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811September 30, 2026
- FDAVITEK 2 AST-N447 TEST KIT Product reference 424620July 15, 2026
- FDAVITEK 2 Gram-negative Susceptibility Cards Containing o Polymyxin B (pb02n) o Ceftazidime/avibactam (cza02n) oFebruary 11, 2026
- FDAVITEK 2 - AST Test kits (in vitro diagnostic test cards for antimicrobial susceptibility): a) AST-N334, REFMay 1, 2024
- FDAbioMerieux Vitek 2 Systems Software version 6.01 software, used with the VITEK 2 automated system. Kit, UPDTAugust 14, 2013
- FDAbioMerieux Vitek 2 Systems Software version 6.01 software, used with the VITEK 2 automated system. Kit, UPDTAugust 14, 2013
Questions about this recall
Was bioMerieux Vitek 2 Systems Software version 6.01 software, used with the VITEK 2 automated system. Kit, UPDT V2S 6.01 DA, SV. Part number: 414567. The product is used with the VITEK 2 automated system which consists of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. recalled?
Yes. FDA published recall Z-1918-2013 on August 14, 2013. Reported hazard: A product issue has been identified for customers using software version 6.01 and are interfaced with a Laboratory Information System (LIS). During the update of software from version 5.xx to 6.01, the Bid-directional Computer interface (BCI) antibiotic codes are not updated properly for Cefovecin, Prulifloxacin and Ceftaroline. For the referenced antibiotics, the firm code in the reference table
What should I do if I have this product?
Voluntary: Firm initiated · Letter