Medtronic Duet External Drainage and Monitoring System, Interlink Injection Sites, Ventricular Catheter, Catalog No. 46915. Intended to allow drainage and measurement of cerebrospinal fluid (CSF) and other fluids of similar physical character from the ventricles or the lumbar subarachnoid (intrathecal) region.

Published July 9, 2014 · Recall initiated: June 9, 2014

Summary

FDA records medical device Class I recall Z-1922-2014, published July 9, 2014 and initiated June 9, 2014. Recalling firm: Medtronic Neurosurgery; Status: Terminated; Reported quantity: 760 units; Lot: Lot No. 206923344 206986678 207096628 207565426 207659574 207945035 207982846 208031201 208055143. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Medtronic Neurosurgery is recalling the Medtronic Duet External Drainage and Monitoring Systems because the patient line tubing in the Duet External Drainage and Monitoring Systems may become disconnected from the patient line stopcock.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Medtronic Neurosurgery
Distributor
Worldwide Distribution -- USA, Australia, Belgium, Canada, Croatia, Cyprus, Denmark, Egypt, Finland, France, Greece, Iran, Israel, Italy, Kuwait, Netherlands, Norway, Poland, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Switzerland, Turkey, and United Kingdom.
Lot
Lot No. 206923344 206986678 207096628 207565426 207659574 207945035 207982846 208031201 208055143
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Medtronic Duet External Drainage and Monitoring System, Interlink Injection Sites, Ventricular Catheter, Catalog No. 46915. Intended to allow drainage and measurement of cerebrospinal fluid (CSF) and other fluids of similar physical character from the ventricles or the lumbar subarachnoid (intrathecal) region. recalled?

Yes. FDA published recall Z-1922-2014 on July 9, 2014. Reported hazard: Medtronic Neurosurgery is recalling the Medtronic Duet External Drainage and Monitoring Systems because the patient line tubing in the Duet External Drainage and Monitoring Systems may become disconnected from the patient line stopcock.

What should I do if I have this product?

Voluntary: Firm initiated · Letter