DISCOVERY ELBOW prosthesis, 4 MM X 100 MM RIGHT HUMERAL FLANGED I BOND COAT Tl 6AL 4V ALLOY. Elbow joint replacement prostheses are intended for primary and revision joint arthroplasty for use in cemented applications.
Published July 2, 2014 · Recall initiated: May 15, 2014
Summary
FDA records medical device Class II recall Z-1925-2014, published July 2, 2014 and initiated May 15, 2014. Recalling firm: Biomet, Inc.; Status: Terminated; Reported quantity: 1; Lot: Catalog/REF #: 114905, Lot #: M846640. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Surface finish is different than specified. The implant has 30 grit blast on the exterior of the flange when the print calls for the glass bead blast. The patient may experience pain requiring surgical intervention.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Biomet, Inc.
- Distributor
- US Nationwide Distribution in the state of California
- Lot
- Catalog/REF #: 114905, Lot #: M846640
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475640 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475620 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475625 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475630 Product Description: Intramedullary BoneAugust 12, 2026
- FDAANSPACH***REF 98-0033***Custom Autolube Food Guard***Rx Only*** This device is intended to be used withJuly 2, 2014
- FDASiemens Dimension LOCI TNI Troponin I Calibrator RC 621. The LOCI TNI Troponin I Calibrator is an in vitroJuly 2, 2014
Questions about this recall
Was DISCOVERY ELBOW prosthesis, 4 MM X 100 MM RIGHT HUMERAL FLANGED I BOND COAT Tl 6AL 4V ALLOY. Elbow joint replacement prostheses are intended for primary and revision joint arthroplasty for use in cemented applications. recalled?
Yes. FDA published recall Z-1925-2014 on July 2, 2014. Reported hazard: Surface finish is different than specified. The implant has 30 grit blast on the exterior of the flange when the print calls for the glass bead blast. The patient may experience pain requiring surgical intervention.
What should I do if I have this product?
Voluntary: Firm initiated · Letter