Magellan Diagnostics LeadCare Ultra Blood Lead Test Kits Part Number: 80-0010

Published May 24, 2017 · Recall initiated: November 29, 2014

Summary

FDA records medical device Class I recall Z-1926-2017, published May 24, 2017 and initiated November 29, 2014. Recalling firm: Magellan Diagnostics, Inc.; Status: Terminated; Reported quantity: 1016; Lot: Kit lots: 1309BU,1310AU,1312BU,1404AU,1408BU. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Magellan Diagnostics, Inc.
Distributor
Nationwide Distribution
Lot
Kit lots: 1309BU,1310AU,1312BU,1404AU,1408BU
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Magellan Diagnostics LeadCare Ultra Blood Lead Test Kits Part Number: 80-0010 recalled?

Yes. FDA published recall Z-1926-2017 on May 24, 2017. Reported hazard: LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.