Microbiologics 217 Osseo Avenue North, St. Cloud, MN 56303, Gram Stain Control Slides, SLO3-10, ATCC Licensed Derivative, MediMart Europe, 11 rue Emile, Zola B.P. 2332, 38933 Grenoble Cedex 2 France. They are microscope slide preparations contains specific organisms populations of known and predictable characteristics. These slide support formal quality assurance programs by serving as quality control challenges to demonstrate and document satisfactory performance of staining regents and methods used in staining procedures and to document personnel's ability to perform the procedure and to properly interpret the microscopic test results.
Published July 11, 2012 · Recall initiated: May 29, 2012
Summary
FDA records medical device Class II recall Z-1936-2012, published July 11, 2012 and initiated May 29, 2012. Recalling firm: Microbiologics Inc; Status: Terminated; Reported quantity: 13,720 slides (10 slides per box); Lot: 0353 - 0355, 0357 - 0363. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Microbiologics is recalling a number of lots of KWIK-QC Gram Stain Slides. This product gram negative control organism (E.coli) is not properly fixated on the slide, which may cause the user to unintentionally contaminate the patient area of the slide by flooding the control well with stain. This contamination could potentially lead to misdiagnosis and improper treatment of a patient.
What to do
Voluntary: Firm initiated · Telephone
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Microbiologics Inc
- Distributor
- Worldwide Distribution-USA (nationwide) including the states of AZ, CA, CO, FL, GA, IL, IA, KY, ME, MA, MI, MN, MS, NJ, NM, NY, PA, PR, SC, TX, VT, and VA. and the countries of SWITZERLAND, ITALY, ISRAEL, IRELAND, COLOMBIA, SWEDEN, INDIA, SINGAPORE, SERBIA, SAUDI ARABIA, CANADA, TAIWAN, SOUTH AFRICA, FRANCE, BRAZIL, MALTA, MACEDONIA, MALAYSIA, FINLAND, ECUADOR, VENEZUELA, REPUBLIC OF KOREA, NETHERLANDS, PERU, UNITED ARAB EMERITES, TURKEY, POLAND, and LEBANON.
- Lot
- 0353 - 0355, 0357 - 0363
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAKWIK-STIK, Quality control kit for culture media, Vibrio parahaemolyticus derived from ATCC 17802, CatalogDecember 10, 2025
- FDALYFO-DISK, M. Canis Derived from ATCC 36299, packaged as: 1) REF 0894K (6-pack), 2) REF 0894P (2-pack)August 6, 2025
- FDAKWIK-STICK, Quality control kit for culture media, Catalog Numbers 01065K, Oligella urethralis derived fromJune 18, 2025
- FDAMicrobiologics Multi-Strain Human Papillomavirus (HPV) Verification Panel (Inactivated Pellet), REF 8235;February 19, 2025
- FDAVacuette Quickshield Complete Plus, 21G x 1 (0.8 x 25 mm), Greiner bio-one. The Quick shield Complete Plus isJuly 11, 2012
- FDADepuy Spine Side Fire Needles 13G X 6IN, part of the Confidence Spinal Cement System Product Code 283904613.July 11, 2012
Questions about this recall
Was Microbiologics 217 Osseo Avenue North, St. Cloud, MN 56303, Gram Stain Control Slides, SLO3-10, ATCC Licensed Derivative, MediMart Europe, 11 rue Emile, Zola B.P. 2332, 38933 Grenoble Cedex 2 France. They are microscope slide preparations contains specific organisms populations of known and predictable characteristics. These slide support formal quality assurance programs by serving as quality control challenges to demonstrate and document satisfactory performance of staining regents and methods used in staining procedures and to document personnel's ability to perform the procedure and to properly interpret the microscopic test results. recalled?
Yes. FDA published recall Z-1936-2012 on July 11, 2012. Reported hazard: Microbiologics is recalling a number of lots of KWIK-QC Gram Stain Slides. This product gram negative control organism (E.coli) is not properly fixated on the slide, which may cause the user to unintentionally contaminate the patient area of the slide by flooding the control well with stain. This contamination could potentially lead to misdiagnosis and improper treatment of a patient.
What should I do if I have this product?
Voluntary: Firm initiated · Telephone