FDAFDA · Food and drugs·Recall number Z-1947-2013

1) Adult Manual Resuscitator, 2K8004, Oxygen Reservoir Bag, Adult Mask; 2) Adult Manual Resuscitator, 2K8005, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask; 3) Adult Manual Resuscitator, 2K8017, Variable Volume Oxygen Reservoir Tubing, Adult Mask; 4) Adult Manual Resuscitator, 2K8034, 40" (1.0 m) Oxygen Reservoir Tubing, PEEP Valve, without Mask; 5) Adult Manual Resuscitator, 2K8035, Oxygen Reservoir Bag, PEEP Valve, Adult Mask; 6) Pediatric Manual Resuscitator, 2K8037, 40" (1.0 m) Oxygen Reservoir Tubing, PEEP Valve, Pediatric Mask, Pressure Limiting Valve with Lock; 7) Infant Manual Resuscitator, 2K8040, Oxygen Reservoir Bag, PEEP Valve, Infant Mask, Pressure Limiting Valve with Lock Product Usage: Pulmonary resuscitation

Published August 21, 2013 · Recall initiated: April 22, 2013

Summary

FDA records medical device Class II recall Z-1947-2013, published August 21, 2013 and initiated April 22, 2013. Recalling firm: Carefusion 2200 Inc; Status: Terminated; Reported quantity: 14,112 units; Lot: 1) Product Code 2K8005; Lot Numbers: 0000355656, 0000355659, 0000355662, 0000355664, 0000355665, 0000355666, 0000355668; 2) Product Code 2K8004; Lot Number: 0000355673; 3) Product Code 2K8017; Lot Numbers: 0000355676, 0000355677; 4) Product Code 2K8034; Lot Number: 0000355688; 5) Product Code 2K8035; Lot Number: 0000358023; 6) Product Code 2K8037; Lot Number: 0000358025; 7) Product Code 2K8040; Lot Number: 0000358026. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

CareFusion has received four (4) customer reports of a component (oxygen connector) within the resuscitation bag being occluded which disables the flow of supplemental oxygen to the resuscitation bag and thus may pose a serious patient safety risk.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Carefusion 2200 Inc
Distributor
US Nationwide in the states of AZ, CA, CO, DE, FL, GA, IL, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NM, NY, OH, PA, TX, UT, WA
Lot
1) Product Code 2K8005; Lot Numbers: 0000355656, 0000355659, 0000355662, 0000355664, 0000355665, 0000355666, 0000355668; 2) Product Code 2K8004; Lot Number: 0000355673; 3) Product Code 2K8017; Lot Numbers: 0000355676, 0000355677; 4) Product Code 2K8034; Lot Number: 0000355688; 5) Product Code 2K8035; Lot Number: 0000358023; 6) Product Code 2K8037; Lot Number: 0000358025; 7) Product Code 2K8040; Lot Number: 0000358026
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was 1) Adult Manual Resuscitator, 2K8004, Oxygen Reservoir Bag, Adult Mask; 2) Adult Manual Resuscitator, 2K8005, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask; 3) Adult Manual Resuscitator, 2K8017, Variable Volume Oxygen Reservoir Tubing, Adult Mask; 4) Adult Manual Resuscitator, 2K8034, 40" (1.0 m) Oxygen Reservoir Tubing, PEEP Valve, without Mask; 5) Adult Manual Resuscitator, 2K8035, Oxygen Reservoir Bag, PEEP Valve, Adult Mask; 6) Pediatric Manual Resuscitator, 2K8037, 40" (1.0 m) Oxygen Reservoir Tubing, PEEP Valve, Pediatric Mask, Pressure Limiting Valve with Lock; 7) Infant Manual Resuscitator, 2K8040, Oxygen Reservoir Bag, PEEP Valve, Infant Mask, Pressure Limiting Valve with Lock Product Usage: Pulmonary resuscitation recalled?

Yes. FDA published recall Z-1947-2013 on August 21, 2013. Reported hazard: CareFusion has received four (4) customer reports of a component (oxygen connector) within the resuscitation bag being occluded which disables the flow of supplemental oxygen to the resuscitation bag and thus may pose a serious patient safety risk.

What should I do if I have this product?

Voluntary: Firm initiated · Letter