Philips Ingenia Ambition S with MR Elastography (MRE). 1. Model Number (REF): 781359. 2. Model Number (REF): 782108. 3. Model Number (REF): 782139.
Published May 6, 2026 · Recall initiated: April 14, 2026
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Philips North America
- Distributor
- Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar, Netherlands, Pakistan, Palestine, Peru, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Kingdom.
- Lot
- 1. Model Number (REF): 781359. UDI: (01)00884838090057(21)48218, (01)00884838090057(21)48233. Serial Numbers: 48218, 48233. 2. Model Number (REF): 782108. UDI: (01)00884838098343(21)48652, (01)00884838098343(21)48551, (01)00884838098343(21)48669. Serial Numbers: 48652, 48551, 48669. 3. Model Number (REF): 782139. UDI: (01)00884838108639(21)29076, (01)00884838108639(21)29079. Serial Numbers: 29076, 29079.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States