FDAFDA · Food and drugs·Recall number Z-1960-2026

Philips Ingenia Ambition S with MR Elastography (MRE). 1. Model Number (REF): 781359. 2. Model Number (REF): 782108. 3. Model Number (REF): 782139.

Published May 6, 2026 · Recall initiated: April 14, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Philips North America
Distributor
Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar, Netherlands, Pakistan, Palestine, Peru, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Kingdom.
Lot
1. Model Number (REF): 781359. UDI: (01)00884838090057(21)48218, (01)00884838090057(21)48233. Serial Numbers: 48218, 48233. 2. Model Number (REF): 782108. UDI: (01)00884838098343(21)48652, (01)00884838098343(21)48551, (01)00884838098343(21)48669. Serial Numbers: 48652, 48551, 48669. 3. Model Number (REF): 782139. UDI: (01)00884838108639(21)29076, (01)00884838108639(21)29079. Serial Numbers: 29076, 29079.
Risk classification
Class II
Area (FDA)
device
Country
United States