EXACTAMIX 1200 Compounder, model numbers 1200-DY, 1200-DX, 1200-DYR and 1200-DXR; EM1200 DY Display, Replacement EM 1200 DY Display. EXACTAMIX 2400 Compounder model numbers 2400-DY, 2400-DX, 2400-DYR and 2400-DYX. Baxa Corporation. An automated pumping system that compounds multiple sterile ingredients into a finished solution in a single patient IV administration bag.
Published July 9, 2014 · Recall initiated: June 10, 2014
Summary
FDA records medical device Class II recall Z-1963-2014, published July 9, 2014 and initiated June 10, 2014. Recalling firm: Baxter Healthcare Corp.; Status: Terminated; Reported quantity: 1758; Lot: model numbers 1200-DY, 1200-DX, 1200-DYR and 1200-DXR; EXACTAMIX 2400 Compounder model numbers 2400-DY, 2400-DX, 2400-DYR and 2400-DYX. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
If the universal ingredient (UI) in an active configuration is changed using the Configuration Editor, a flush of the outlet pump tube will not be initiated by the software. It could result in the original UI remaining in the tube which could be delivered into the next bag.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Baxter Healthcare Corp.
- Distributor
- US Nationwide
- Lot
- model numbers 1200-DY, 1200-DX, 1200-DYR and 1200-DXR; EXACTAMIX 2400 Compounder model numbers 2400-DY, 2400-DX, 2400-DYR and 2400-DYX
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGAMBRO Capillary Dialyzer, Polyflux(R) Revaclear MAX, STERILE, Product code 110634 Product Usage: TheNovember 16, 2016
- FDASynovis VASCU-GUARD Peripheral Vascular Patch in the following sizes and product codes: 1 cm x 6 cm, productAugust 24, 2016
- FDA0.9% Sodium Chloride Injection, USP, 250 mL VIAFLEX Plastic Container Bag, Rx Only, Manufactured by BaxterJune 15, 2016
- FDADextrose Injection,USP,70% 2000 mL VIAFLEX Plastic Container Bag, Pharmacy Bulk Package, Not For DirectJune 15, 2016
- FDAACUSON S Family Ultrasound Systems at software versions VC25D, VC30A and VC30B; (S2000, S3000 and S1000July 9, 2014
- FDASystem Manual for the da Vinci Standard Surgical System; da Vinci S System and da Vinci Si System. IntuitiveJuly 9, 2014
Questions about this recall
Was EXACTAMIX 1200 Compounder, model numbers 1200-DY, 1200-DX, 1200-DYR and 1200-DXR; EM1200 DY Display, Replacement EM 1200 DY Display. EXACTAMIX 2400 Compounder model numbers 2400-DY, 2400-DX, 2400-DYR and 2400-DYX. Baxa Corporation. An automated pumping system that compounds multiple sterile ingredients into a finished solution in a single patient IV administration bag. recalled?
Yes. FDA published recall Z-1963-2014 on July 9, 2014. Reported hazard: If the universal ingredient (UI) in an active configuration is changed using the Configuration Editor, a flush of the outlet pump tube will not be initiated by the software. It could result in the original UI remaining in the tube which could be delivered into the next bag.
What should I do if I have this product?
Voluntary: Firm initiated · Letter