Vaxcel PICC with PASV Introducer Kit under the following labels: 1) 4F, UPN M001454560, Rx ONLY & 2) 5F, UPN M001454610, Rx ONLY. The Vaxcel PICC with PASV is indicated for use in establishing peripheral access to the central venous system for administration of fluids including, but not limited to, hydration agents, antibiotics, chemotherapy, analgesics, nutritional therapy, and blood products. It is also indicated for blood specimen withdrawal. This product is intended for central venous access in adults, children and infants who require intravenous (IV) therapy.
Published June 22, 2016 · Recall initiated: March 14, 2016
Summary
FDA records medical device Class II recall Z-1967-2016, published June 22, 2016 and initiated March 14, 2016. Recalling firm: Navilyst Medical, Inc., an AngioDyamics Company; Status: Terminated; Reported quantity: 328 units; Lot: Batch/Lots:1) 4714745 with Use By 2016-03-31. Batch/Lots: 2) 4745413, 4819467, 4827954, 4725425 & 4726948 with Use By Date Range 2016-03-31 to 2016-11-30. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
PICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Navilyst Medical, Inc., an AngioDyamics Company
- Distributor
- Worldwide Distribution US (Nationwide) Internationally to AU, GR, SE, SA, IN, CA, GB, BR, BE, NL, ES, KR, HK, AE, IT and ES.
- Lot
- Batch/Lots:1) 4714745 with Use By 2016-03-31. Batch/Lots: 2) 4745413, 4819467, 4827954, 4725425 & 4726948 with Use By Date Range 2016-03-31 to 2016-11-30.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVaxcel PICC with PASV Catheter Kit under the following labels: 1) 3F, UPN H965454350, Rx ONLY, 2) 4F, UPNJune 22, 2016
- FDABioFlo PICC with ENDEXO and PASV Valve Technology, MST-45 Kit, 6F-55cm, UPN H965458440 , Rx ONLY The XcelaJune 22, 2016
- FDABioFlo PICC with ENDEXO and PASV Valve Technology, Catheter Kit, 6F-55cm, UPN H965458410, Rx ONLY The XcelaJune 22, 2016
- FDABioSync Anatomic Cotton Wedge, 16 x 6.5 mm.; Used for internal bone fixation for bone fractures, fusions orJune 22, 2016
- FDABioSync Anatomic Cotton Wedge, 16 x 5.5 mm.; Used for internal bone fixation for bone fractures, fusions orJune 22, 2016
- FDAStryker PowerPRO (XT, IT, or TL) Cots Stretcher, Wheeled The Stryker Power-PRO" are powered ambulance cots areJune 22, 2016
Questions about this recall
Was Vaxcel PICC with PASV Introducer Kit under the following labels: 1) 4F, UPN M001454560, Rx ONLY & 2) 5F, UPN M001454610, Rx ONLY. The Vaxcel PICC with PASV is indicated for use in establishing peripheral access to the central venous system for administration of fluids including, but not limited to, hydration agents, antibiotics, chemotherapy, analgesics, nutritional therapy, and blood products. It is also indicated for blood specimen withdrawal. This product is intended for central venous access in adults, children and infants who require intravenous (IV) therapy. recalled?
Yes. FDA published recall Z-1967-2016 on June 22, 2016. Reported hazard: PICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
What should I do if I have this product?
Voluntary: Firm initiated · Letter