enFlow IV Fluid Warmer (510k name: EFLOW IV FLUID WARMER, MODEL 100, EFLOW DISPOSABLE CARTRIDGE, MODEL 200 AND EFLOW AC POWER SUPPLY, MODEL 120). The Enginivity enFlow TM IV Fluid Warmer is intended for warming blood, blood products and intravenous solutions prior to administration. It is intended to be used by healthcare professionals in hospital, clinical and field environments to help prevent hypothermia.
Published July 18, 2012 · Recall initiated: February 20, 2012
Summary
FDA records medical device Class II recall Z-1972-2012, published July 18, 2012 and initiated February 20, 2012. Recalling firm: Vital Signs Devices, a GE Healthcare Company; Status: Terminated; Reported quantity: 11803 cases; Lot: a) 510(k): K060537 b) Medical device listing number: D071942 All enFlow Warmer Straps in the field present a potential for tissue injury. Sold separately as Product #: 980304 and 980304EU Included as an accessory in following enFlow products: Military Domestic 91000153 980100 91000154 980100EU 91000156 980105VS 91000103. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The EnFlow Warmer Strap does not meet the requirements of the biocompatibility standard (ISO 10993) for products that may have contact with skin for less than 24 hours.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Vital Signs Devices, a GE Healthcare Company
- Distributor
- Nationwide Distribution.
- Lot
- a) 510(k): K060537 b) Medical device listing number: D071942 All enFlow Warmer Straps in the field present a potential for tissue injury. Sold separately as Product #: 980304 and 980304EU Included as an accessory in following enFlow products: Military Domestic 91000153 980100 91000154 980100EU 91000156 980105VS 91000103
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADisposable Temperature Probes, (sterile), Sensor diameter: 3mm/0.12 in; 25 units per box. Vital Signs Inc. aAugust 21, 2013
- FDACorometrics Qwik Connect Plus Spiral Electrode- Model #7000AA0 A fetal scalp circular (spiral) electrode andNovember 28, 2012
- FDASystem 1 Base 220/240V. The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on theJuly 18, 2012
- FDAiStat1 Analyzer Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree tests for sodium,July 18, 2012
- FDASystem 1 Base 100/120V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on theJuly 18, 2012
- FDAi-Stat Portable Clinical Analyzer; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree testsJuly 18, 2012
Questions about this recall
Was enFlow IV Fluid Warmer (510k name: EFLOW IV FLUID WARMER, MODEL 100, EFLOW DISPOSABLE CARTRIDGE, MODEL 200 AND EFLOW AC POWER SUPPLY, MODEL 120). The Enginivity enFlow TM IV Fluid Warmer is intended for warming blood, blood products and intravenous solutions prior to administration. It is intended to be used by healthcare professionals in hospital, clinical and field environments to help prevent hypothermia. recalled?
Yes. FDA published recall Z-1972-2012 on July 18, 2012. Reported hazard: The EnFlow Warmer Strap does not meet the requirements of the biocompatibility standard (ISO 10993) for products that may have contact with skin for less than 24 hours.
What should I do if I have this product?
Voluntary: Firm initiated · Letter