CNS-6200 Series Central Nurse Station and accessories. Model Number: CNS-6201 For cardiac and vital signs monitoring for multiple patients.
Published July 9, 2014 · Recall initiated: June 3, 2014
Summary
FDA records medical device Class II recall Z-1979-2014, published July 9, 2014 and initiated June 3, 2014. Recalling firm: Nihon Kohden America Inc; Status: Terminated; Reported quantity: 430; Lot: Model Number: CNS-6201 Software Version 02-26 only. Serial Numbers: 00485, 00488, 00489, 00490, 00491, 00361, 00429, 00430, 00432, 00436, 00443, 00447, 00478, 00486, 00546, 00564, 00573, 00574, 00575, 00576, 00592, 00626, 00638, 00643, 00651, 00674, 00675, 00677, 00678, 00679, 00687. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Software Version 02-26, when used with the Central Nursing Station (CNS) 6201, (PU-621 RA) may unexpectedly and without warning reboot, resulting in a period of approximately 3 minutes during which time the system does not display patient data and does not emit alarms based on that data should they be appropriate.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Nihon Kohden America Inc
- Distributor
- Worldwide Distribution - US in the state of Oklahoma and in the countries of Bolivia, Chile, Germany, Japan, Korea, Taiwan, Yemen, and South Africa.
- Lot
- Model Number: CNS-6201 Software Version 02-26 only. Serial Numbers: 00485, 00488, 00489, 00490, 00491, 00361, 00429, 00430, 00432, 00436, 00443, 00447, 00478, 00486, 00546, 00564, 00573, 00574, 00575, 00576, 00592, 00626, 00638, 00643, 00651, 00674, 00675, 00677, 00678, 00679, 00687.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAdult Cap-ONE Biteblock REF YG-227T The cap-ONE Biteblock is intended to sample CO2 and administerNovember 13, 2024
- FDANihon Kohden Adult Ear Clip SpO2 Sensor, 1.5 meter Model 809030007 Reusable SpO2 Ear Clip, eachNovember 6, 2024
- FDANihon Kohden Adult/Pediatric Forehead Disposable SpO2 Sensor, 0.9m Model 809030006 Disp. SpO2 Forehead Probe,November 6, 2024
- FDABSM-3000 Series Bedside Monitor REF BSM-3572A The Life Scope BSM-3000 Series Bedside Monitoring System isJune 12, 2024
- FDAStryker Orthopaedics Triathlon Tritanium Patella Inserter, . NON-STERILE Product Usage: The TriathlonJuly 9, 2014
- FDANon-Mydriatic Auto Fundus Camera. Model AFC-330. An ophthalmic camera for use in capturing images of theJuly 9, 2014
Questions about this recall
Was CNS-6200 Series Central Nurse Station and accessories. Model Number: CNS-6201 For cardiac and vital signs monitoring for multiple patients. recalled?
Yes. FDA published recall Z-1979-2014 on July 9, 2014. Reported hazard: Software Version 02-26, when used with the Central Nursing Station (CNS) 6201, (PU-621 RA) may unexpectedly and without warning reboot, resulting in a period of approximately 3 minutes during which time the system does not display patient data and does not emit alarms based on that data should they be appropriate.
What should I do if I have this product?
Voluntary: Firm initiated · Letter