Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and PET/CT systems: Model - MX8000 Models: 728120/728125/728201

Published July 7, 2021 · Recall initiated: May 1, 2021

Summary

FDA records medical device Class II recall Z-1979-2021, published July 7, 2021 and initiated May 1, 2021. Recalling firm: Philips North America Llc; Status: Ongoing; Reported quantity: 1,301 Monitor Bracket Assemblies total; Lot: Model/ Serial Number/UDI: 728120 2516 N/A 728120 3179 N/A 728120 3316 N/A 728201 9326 N/A 728120 3304 N/A 728120 3271 N/A 728120 2537 N/A 728120 3324 N/A 728120 H3321 N/A 728120 3058 N/A 728125 2529 N/A. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips North America Llc
Distributor
Worldwide distribution - US Nationwide and the countries of Canada, Mexico, AE, AR, AT, AU, BD, BE, BR, CH, CL, CN, CO, CU, CY, CZ, DE, DK, DZ, EE, EG, ES, FI, FR, GB, GE, GI, GL, GR, HK, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KG, KH, KR, LA, LB, LK, LT, LY, MM, MN, MU, MV, MY, NI, NL, NO, NZ, OM, PA, PF, PH, PK, PL, PS, PT, QA, RE, RU, RW, SA, SE, SG , SI, SK, TH, TN, TR, UZ, VN, XK, YE, ZA, ZW.
Lot
Model/ Serial Number/UDI: 728120 2516 N/A 728120 3179 N/A 728120 3316 N/A 728201 9326 N/A 728120 3304 N/A 728120 3271 N/A 728120 2537 N/A 728120 3324 N/A 728120 H3321 N/A 728120 3058 N/A 728125 2529 N/A
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and PET/CT systems: Model - MX8000 Models: 728120/728125/728201 recalled?

Yes. FDA published recall Z-1979-2021 on July 7, 2021. Reported hazard: Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.

What should I do if I have this product?

Voluntary: Firm initiated · Letter