GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S

Published June 25, 2025 · Recall initiated: May 8, 2025

Summary

FDA records medical device Class II recall Z-1980-2025, published June 25, 2025 and initiated May 8, 2025. Recalling firm: Ohio Medical Corporation; Status: Ongoing; Reported quantity: 36 units; Lot: Serial Numbers: ZL316367, ZL316446, ZL316462, ZL316475, ZL316481, ZL316484, ZL316490, ZL316493, ZL316503, ZL317057, ZL317087, ZL317088, ZL317090, ZL322386, ZL322387, ZL322390, ZL322392, ZL322409, ZL322412, ZL322439, ZL322441, ZL322442, ZL322443, ZL322445, ZL322447, ZL322451, ZL322455, ZL322456, ZL322462, ZL322463, ZL322466, ZL322467, ZL322470, ZL322472, ZL322473, ZL322837. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Erroneous calibration values, potentially resulting in: as the device ages, oxygen purity will gradually decrease to the point where the device alarms to indicate low oxygen purity will be incorrect.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ohio Medical Corporation
Distributor
US Nationwide distribution in the states of AR, FL, VA, ND, MS, MN, CA, NV, KY.
Lot
Serial Numbers: ZL316367, ZL316446, ZL316462, ZL316475, ZL316481, ZL316484, ZL316490, ZL316493, ZL316503, ZL317057, ZL317087, ZL317088, ZL317090, ZL322386, ZL322387, ZL322390, ZL322392, ZL322409, ZL322412, ZL322439, ZL322441, ZL322442, ZL322443, ZL322445, ZL322447, ZL322451, ZL322455, ZL322456, ZL322462, ZL322463, ZL322466, ZL322467, ZL322470, ZL322472, ZL322473, ZL322837.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S recalled?

Yes. FDA published recall Z-1980-2025 on June 25, 2025. Reported hazard: Erroneous calibration values, potentially resulting in: as the device ages, oxygen purity will gradually decrease to the point where the device alarms to indicate low oxygen purity will be incorrect.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail