Dover 100% Silicone Silver-Hydrogel Coated Foley Catheter with Temperature Sensor 16FR 5CCCC2W, Product Code: 50516ICT - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
Published May 20, 2020 · Recall initiated: April 30, 2020
Summary
FDA records medical device Class II recall Z-1985-2020, published May 20, 2020 and initiated April 30, 2020. Recalling firm: Covidien Llc; Status: Terminated; Reported quantity: 480 units; Lot: Lot Number: 1909500464. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2®C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Covidien Llc
- Distributor
- US Nationwide distribution.
- Lot
- Lot Number: 1909500464
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPalindrome Precision HSI Chronic Catheter, Sterile, Single Use, Rx Only. For acute and chronic hemodialysis,July 13, 2022
- FDAPalindrome Precision Chronic Catheter, Sterile, Single Use, Rx Only. For vascular access for acute and chronicJuly 13, 2022
- FDAPuritan Bennett 980 Series Ventilator, 980U1ESDIUCDecember 15, 2021
- FDAPuritan Bennett 980 Series Ventilator, 980X2ESDIECDecember 15, 2021
- FDABD PosiFlush SF Saline Flush Syringe 10 mL used in POWERWAND XL 5Fr 8cm Max Barrier Kit. REF/UDI /May 20, 2020
- FDADover 100% Silicone Premium Urine Meter Foley Tray, 400mL, 14 Fr/Ch (4.7 mm), 5 mL Catheter Pre-connected toMay 20, 2020
Questions about this recall
Was Dover 100% Silicone Silver-Hydrogel Coated Foley Catheter with Temperature Sensor 16FR 5CCCC2W, Product Code: 50516ICT - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only. recalled?
Yes. FDA published recall Z-1985-2020 on May 20, 2020. Reported hazard: Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2®C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
What should I do if I have this product?
Voluntary: Firm initiated · Letter