Back to search
FDAFDA · Food and drugs·Recall number Z-1998-2012

Bayonet Forceps 7.-1/8 (181.0mm) 0.75mm Non-Insulated Tips, Sterile Single Use Surgical/Bipolar Forceps, Model #20-1360K 510 (k) #K884656. The surgical device is package one (s) per plastic Tyvek peel pouch, 10 plastic Tyvek peel pouches per box. Bipolar Forceps and Cords - Olsen Medical Bipolar Forceps and Cords are intended for use in coagulating tissue during surgical procedures.

Published July 18, 2012 · Recall initiated: June 1, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Firm officials reported to CIN-DO that for this device, in the labeling on the inner Tyvek peel pouch, the letter 'E' in Electrosurgical was turned into an 'X' and the number '5" of the (mm tips) was turned into a '0'. This affected the description and some of the text of the labels. However, it did not t affect the Part Number and Lot Number on the label.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
PSC Industries Inc
Distributor
Worldwide distribution: USA (nationwide) including states of: CA, CT, FL, GA, MD, MI, NC, OR, SC, and TX and countries of: Canada and Philippines.
Lot
Model/part Number: 20-1360K, Lot Number: 023420,, Expiration Date: 2/28/2017.
Risk classification
Class II
Area (FDA)
device
Country
United States