FDAFDA · Food and drugs·Recall number Z-1998-2012
Bayonet Forceps 7.-1/8 (181.0mm) 0.75mm Non-Insulated Tips, Sterile Single Use Surgical/Bipolar Forceps, Model #20-1360K 510 (k) #K884656. The surgical device is package one (s) per plastic Tyvek peel pouch, 10 plastic Tyvek peel pouches per box. Bipolar Forceps and Cords - Olsen Medical Bipolar Forceps and Cords are intended for use in coagulating tissue during surgical procedures.
Published July 18, 2012 · Recall initiated: June 1, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Firm officials reported to CIN-DO that for this device, in the labeling on the inner Tyvek peel pouch, the letter 'E' in Electrosurgical was turned into an 'X' and the number '5" of the (mm tips) was turned into a '0'. This affected the description and some of the text of the labels. However, it did not t affect the Part Number and Lot Number on the label.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- PSC Industries Inc
- Distributor
- Worldwide distribution: USA (nationwide) including states of: CA, CT, FL, GA, MD, MI, NC, OR, SC, and TX and countries of: Canada and Philippines.
- Lot
- Model/part Number: 20-1360K, Lot Number: 023420,, Expiration Date: 2/28/2017.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States