00597001303 Femur CR PRC FEM COMP SIZE CML Rx, Sterile; 00597001304 Femur CR PRC FEM COMP SIZE CMR Rx, Sterile; 00597001403 Femur CR PRC FEM COMP SIZE DML Rx, Sterile; 00597001404 Femur CR PRC FEM COMP SIZE DMR Rx, Sterile; 00597001503 Femur CR PRC FEM COMP SIZE EML Rx, Sterile; 00597001504 Femur CR PRC FEM COMP SIZE EMR Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Published August 28, 2013 · Recall initiated: June 7, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Worldwide Distribution-USA (nationwide) and the country of Japan.
- Lot
- all codes
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.