Back to search
FDAFDA · Food and drugs·Recall number Z-2010-2016

BIOGRAPH 40-3R to 64-3R Upgrade, Material Number 10246390 The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

Published June 29, 2016 · Recall initiated: April 22, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Medical Solutions USA Inc.
Distributor
Worldwide Distribution - US Nationwide
Lot
Serial Number 1001
Risk classification
Class II
Area (FDA)
device
Country
United States