Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. OSC COLEMAN FAT HARVEST KIT, Medline Kit Number/SKU DYNJ32854B; 2. RR-DR. CHHETRI MDL PACK, Medline Kit Number/SKU DYNJ62635B.
Published May 6, 2026 · Recall initiated: March 19, 2026
Summary
FDA records medical device Class II recall Z-2018-2026, published May 6, 2026 and initiated March 19, 2026. Recalling firm: Medline Industries, LP; Status: Ongoing; Reported quantity: 208 kits; Lot: Medline Kit Number/SKU DYNJ32854B: UDI/DI each 10198459413414, UDI/DI case 40198459413415, Lot Number : 25GBU656; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25GMD214; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25FMC376; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25DMJ030; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25BME902; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25AMJ032; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 24IMH612; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 24EME486; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 24DMD315; Medline Kit Number/SKU DYNJ62635C: UDI/DI each 10198459485619, UDI/DI case 40198459485610, Lot Number : 26BMD607; Medline Kit Number/SKU DYNJ62635C: UDI/DI each 10198459485619, UDI/DI case 40198459485610, Lot Number : 25JMD196. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medline Industries, LP
- Distributor
- Worldwide distribution.
- Lot
- Medline Kit Number/SKU DYNJ32854B: UDI/DI each 10198459413414, UDI/DI case 40198459413415, Lot Number : 25GBU656; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25GMD214; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25FMC376; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25DMJ030; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25BME902; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25AMJ032; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 24IMH612; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 24EME486; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 24DMD315; Medline Kit Number/SKU DYNJ62635C: UDI/DI each 10198459485619, UDI/DI case 40198459485610, Lot Number : 26BMD607; Medline Kit Number/SKU DYNJ62635C: UDI/DI each 10198459485619, UDI/DI case 40198459485610, Lot Number : 25JMD196.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMEDLINE medical convenience kits labeled as: 1) HIP SCOPE PACK, Medline Kit Number/SKU DYNJ64103C; 2) HIPOctober 28, 2026
- FDAMEDLINE medical convenience kits labeled as: 1) BIRTHDAY KIT-DOUBLE BASIN-LF, Medline Kit Number/SKUOctober 28, 2026
- FDAMEDLINE DRAPE C-SECTION FEN CLR SCREEN W, Medline Product Number/SKU DYNJP6108October 28, 2026
- FDAMEDLINE medical convenience kits labeled as: FETOSCOPY CDS-LF, Medline Product Number/SKU CDS840215POctober 28, 2026
- FDAMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. VP SHUNT, Medline KitMay 6, 2026
- FDAMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. RR-OMFS PACK-LF,May 6, 2026
Questions about this recall
Was Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. OSC COLEMAN FAT HARVEST KIT, Medline Kit Number/SKU DYNJ32854B; 2. RR-DR. CHHETRI MDL PACK, Medline Kit Number/SKU DYNJ62635B. recalled?
Yes. FDA published recall Z-2018-2026 on May 6, 2026. Reported hazard: During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit