BD SARS-CoV-2 Reagents for BD MAX" Systems. RT-PCR test to detect SARS-CoV-2 in nasal, nasopharyngeal, oropharyngeal swab specimens, and saliva specimens.

Published June 12, 2024 · Recall initiated: May 7, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Assays may produce false negative results due to decreased activity of the enzyme utilized in manufacturing.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
Becton Dickinson & Co.
Distributor
Domestic: Nationwide Distribution; Foreign: AU, BR, BE, CA, CO, & JP.
Lot
Catalog No. 445003-01; UDI-DI: (01) 60382904450030 (17) 240803 (10) 3291356 (20) 01 (30) 1, (01) 60382904450030 (17) 241119 (10) 3291358 (20) 01 (30) 1, (01) 60382904450030 (17) 241217 (10) 3326001 (20) 01 (30) 1; Lot No. 3291356 (Exp. 8/3/24), 3291358 (Exp. 11/19/24), 3326001 (Exp. 12/17/24).
Risk classification
Class II
Area (FDA)
device
Country
United States