SYS IVK, Bio mCT-S(40) 3R->4R Upgrade, MATERIAL NUMBER 10250743 The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

Published June 29, 2016 · Recall initiated: April 22, 2016

Summary

FDA records medical device Class II recall Z-2035-2016, published June 29, 2016 and initiated April 22, 2016. Recalling firm: Siemens Medical Solutions USA Inc.; Status: Terminated; Reported quantity: 4 units; Lot: Serial Numbers: 1002,1001,11003,11004. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Medical Solutions USA Inc.
Distributor
Worldwide Distribution - US Nationwide
Lot
Serial Numbers: 1002,1001,11003,11004
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was SYS IVK, Bio mCT-S(40) 3R->4R Upgrade, MATERIAL NUMBER 10250743 The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information. recalled?

Yes. FDA published recall Z-2035-2016 on June 29, 2016. Reported hazard: Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.

What should I do if I have this product?

Voluntary: Firm initiated · Letter