Terumo Advanced Perfusion System 1. System 1 base 220/240V. The Terumo Advanced Perfusion System I is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only When used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.
Published July 25, 2012 · Recall initiated: June 18, 2012
Summary
FDA records medical device Class II recall Z-2043-2012, published July 25, 2012 and initiated June 18, 2012. Recalling firm: Terumo Cardiovascular Systems Corporation; Status: Terminated; Reported quantity: 648 units; Lot: Catalog Number 801764 and serial number 0006-1441. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Terumo Cardiovascular Systems has received several reports of malfunctions of the Air Bubble Detection System used with its heart-lung machines over the last three years. The malfunctions were expressed in various failure modes (oversensitivity, false alarms and were found to have various root causes. The malfunctions occurred in systems that were excluded from prior corrective actions or in syste
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Terumo Cardiovascular Systems Corporation
- Distributor
- Worldwide Distribution-USA (nationwide) and the countries of ARGENTINA, BANGLADESH, BELGIUM, Bolivia, BRAZIL, BRISBANE, CANADA, CHILE, China, COLOMBIA, COLOMBIA, Costa Rica, DENMARK, Dominican Republic, Ecuador, Egypt, FRANCE, GERMANY, Guatemala, HOLLAND, Honduras, Hong Kong, India, INDIA, Indonesia, Israel, ITALY, Jordan, KOREA, Kuwait, Lebanon, LEICESTERSHIRE, Malaysia, Mexico, NETHERLANDS, New Zealand, Nicaragua, NORWAY, Pakistan, Panama, PERU, Philippines, Russia, Saudi Arabia, SINGAPORE , SOUTH AFRICA, South Korea, SPAIN, Sri Lanka, SWEDEN, SWITZERLAND, Taiwan, THAILAND, Turkey, UNITED ARAB EMIRATES (UAE), Uruguay, Venezuela, Vietnam, WEST GERMANY and Yemen.
- Lot
- Catalog Number 801764 and serial number 0006-1441
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPMJuly 2, 2025
- FDACardiovascular Procedure Kits that include High Crack Check Valve PN 205307A Part Number: 6375 16010 62974April 10, 2024
- FDA(1) Capiox FX25 (West) Advance Oxygenator- with left port, 4-liter Reservoir, XC modified to Advance, SterileDecember 23, 2020
- FDATerumo Capiox NX19 Oxygenator With Reservoir (East Orientation) P/N: 3CX*NX19REAugust 7, 2019
- FDATerumo Advanced Perfusion System 1. System 1 base 100/120V. The Terumo Advanced Perfusion System I isJuly 25, 2012
- FDASarns Modular Perfusion System 8000 base, 4 pump 115V. The Sarns Modular Perfusion system 8000 is indicatedJuly 25, 2012
Questions about this recall
Was Terumo Advanced Perfusion System 1. System 1 base 220/240V. The Terumo Advanced Perfusion System I is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only When used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment. recalled?
Yes. FDA published recall Z-2043-2012 on July 25, 2012. Reported hazard: Terumo Cardiovascular Systems has received several reports of malfunctions of the Air Bubble Detection System used with its heart-lung machines over the last three years. The malfunctions were expressed in various failure modes (oversensitivity, false alarms and were found to have various root causes. The malfunctions occurred in systems that were excluded from prior corrective actions or in syste
What should I do if I have this product?
Voluntary: Firm initiated · Letter