nordicBrainEX Product Usage: nordicBrainEx is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf PC workstation and can be used to perform image viewing, processing and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities.
Published June 6, 2018 · Recall initiated: December 21, 2012
Summary
FDA records medical device Class II recall Z-2045-2018, published June 6, 2018 and initiated December 21, 2012. Recalling firm: NordicNeuroLab AS; Status: Terminated; Reported quantity: 36 licenses; Lot: version 1.1.2. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is an error related to the relative geometry between fiber tracts in a fiber group and images. The error occurs in the following situations: A group of fibers has been selected using the VOI functionality and put into a fiber group. This is done with images series A visualized in the 3D viewer. Then a new images series B is visualized in the 3D viewer. If image series A and B have the same geometrical resolution (pixel sizes and slice distances), the coregistration matrix taking A to B will not be applied to the fiber group, and therefore the fiber group will in general not be positioned correctly on B in the 3D viewer. Furthermore, if the fiber group is exported as a new image series the fibers could be misplaced, and similarly if the fibers are visualized in the MPR on B, the positioning will not be correct. The misalignment will be equal to the rotations/translations necessary to align A with B.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- NordicNeuroLab AS
- Distributor
- Worldwide Distribution - US Nationwide - Saudi Arabia, Brazil, Sweden, Spain, Russia, Belgium, Norway, South Korea, Japan
- Lot
- version 1.1.2
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Norway
Related recalls
- FDANystatin Oral Suspension, USP, 500,000 units/5 mL, For Institutional Use Only, packaged in 5 mL cups, Rx Only,June 13, 2018
- FDAValsartan Tablets USP, 320 mg, 90 tablet, Rx Only, Manufactured by: Jubilant Generics Ltd. Roorkee - 247661,June 13, 2018
- FDACode - Red Pepper Seasoning (21 oz.) Net Weight: 21 oz., Number of Units per Package: 6 Package Type: clear,June 13, 2018
- FDAVPS Rhythm ECG Accessory Pack with Johans Adapter; Product codes: RHY-100-ECGJJune 6, 2018
- FDAPeel Away Sheath over Dilator; Product Codes: PL-05041 and PL-05052June 6, 2018
- FDA8Fr. Catheter Clamp with Fastener; Product Code: CC-00008June 6, 2018
Questions about this recall
Was nordicBrainEX Product Usage: nordicBrainEx is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf PC workstation and can be used to perform image viewing, processing and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities. recalled?
Yes. FDA published recall Z-2045-2018 on June 6, 2018. Reported hazard: There is an error related to the relative geometry between fiber tracts in a fiber group and images. The error occurs in the following situations: A group of fibers has been selected using the VOI functionality and put into a fiber group. This is done with images series A visualized in the 3D viewer. Then a new images series B is visualized in the 3D viewer. If image series A and B have the same geometrical resolution (pixel sizes and slice distances), the coregistration matrix taking A to B will not be applied to the fiber group, and therefore the fiber group will in general not be positioned correctly on B in the 3D viewer. Furthermore, if the fiber group is exported as a new image series the fibers could be misplaced, and similarly if the fibers are visualized in the MPR on B, the positioning will not be correct. The misalignment will be equal to the rotations/translations necessary to align A with B.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail